Original Article |

Artificial Muscle for Reanimation of the Paralyzed Face Durability and Biocompatibility in a Gerbil Model FREE

Levi G. Ledgerwood, MD; Steven Tinling, PhD; Craig Senders, MD; Annjoe Wong-Foy, PhD; Harsha Prahlad, PhD; Travis T. Tollefson, MD, MPH
[+] Author Affiliations

Author Affiliations: Department of Otolaryngology–Head and Neck Surgery, University of California, Davis Medical Center, Sacramento (Drs Ledgerwood, Tinling, Senders, and Tollefson); and SRI International, Menlo Park, California (Drs Wong-Foy and Prahlad).


Arch Facial Plast Surg. 2012;14(6):413-418. doi:10.1001/archfacial.2012.696.
Text Size: A A A
Published online

Background Current management of permanent facial paralysis centers on nerve grafting and muscle transfer; however, limitations of those procedures call for other options.

Objectives To determine the durability and biocompatibility of implanted artificial muscle in a gerbil model and the degree of inflammation and fibrosis at the host tissue–artificial muscle interface.

Methods Electroactive polymer artificial muscle (EPAM) devices engineered in medical-grade silicone were implanted subcutaneously in 13 gerbils. The implanted units were stimulated with 1 kV at 1 Hz, 24 h/d via a function generator. Electrical signal input/output was recorded up to 40 days after implantation. The animals were euthanized between 23 and 65 days after implantation, and the host tissue–implant interface was evaluated histologically.

Results The animals tolerated implantation of the EPAM devices well, with no perioperative deaths. The muscle devices created motion for a mean of 30.3 days (range, 19-40 days), with a mean of 2.6 × 106 cycles (range, 1.6 × 106 to 3.5 × 106 cycles). Histologic examination of the explanted devices revealed the development of a minimal fibrous capsule surrounding the implants, with no evidence of bacterial infection or inflammatory infiltrate. No evidence of device compromise, corrosion, or silicone breakdown was noted.

Conclusions Artificial muscle implanted in this short-term animal model was safe and functional in this preliminary study. We believe that EPAM devices will be a safe and viable option for restoration of facial motions in patients with irreversible facial paralysis.

Figures in this Article

Paralysis of the facial nerve and associated sequelae affect more than 40 000 patients in the United States each year.1 Most of these cases are self-limiting Bell's palsy; patients with permanent paralysis require surgical rehabilitation. Patients experience psychologically debilitating disfigurement and impaired communication ability, as well as functional impairments, such as oral incompetence to food and saliva, the inability to smile, brow ptosis, nasal obstruction, and eyelid complaints, such as extropion and eye irritation (exposure keratopathy). Chronic facial paralysis often requires surgical management. Corneal protection is paramount to avoid exposure keratopathy and visual impairment. In the acute setting, eye lubrication regimens are used. Paralytic eyelid rehabilitation may involve gold weight eyelid loading and lower eyelid tarsorrhaphy or canthoplasty. The drawbacks of these procedures include potential gold weight extrusion, eyelid ptosis, and creation of dyssynchronous slow, gravity-assisted eyelid closure.2 The potential for using an artificial muscle to recreate facial motion in the setting of permanent facial paralysis may restore facial movement, including eyelid blinking.

Traditional rehabilitation in patients with facial paralysis involves static and dynamic surgical procedures that seek to mimic the function of the paralyzed facial musculature.3 Nerve grafting, when possible, can restore tone and function but can be limited by synkinetic facial movement and spasms. In the setting of permanent facial paralysis, the state of the science has evolved to free microvascular muscle transfer to create facial motion. Significant advances in the technique and outcomes of smile restoration have occurred since Harii et al4 first described the gracilis muscle transfer for facial reanimation in 1976. The motor innervation of a free muscle transfer can include a cross-face nerve graft or grafting to the hypoglossal or masseteric nerve. Bilateral free muscle transplant of the gracilis muscle has become the most popular method to improve function of the lip in patients with Möbius syndrome.49 Free tissue transfer represents a more dynamic solution to facial paralysis than the previous methods; however, these procedures are not without potentially significant complications. Methods of animation with less morbidity would be of great benefit to this population and could be used when free tissue transfer is contraindicated.

Electroactive polymer artificial muscle (EPAM) was developed in the 1990s (SRI International) for several robotic applications. Artificial muscle products function by converting electrical force to mechanical motion. Thin films of dielectric elastomers, such as silicone (20-60 μm thick), are laminated with conductive carbon particles suspended in a soft polymer matrix. When the laminated films are exposed to high-voltage electric fields with low amperage, positive and negative charges accumulate on the opposite electrodes and are attracted to each other, squeezing the polymer insulator, which responds by expanding perpendicularly to the electric field lines; contraction occurs when the electric charge is again removed.10 Dielectric elastomers can grow by as much as 300% of their nonactivated size, creating an efficient, powerful artificial muscle. In general, greater voltage results in greater expansion until electrical breakdown. In addition, the EPAM devices can operate at a very low current, which creates little heat or noise.11

Potential applications for artificial muscle in humans have been theorized for movement of fingers and hands, artificial sphincters, and facial reanimation.12,13 Of these concepts, to our knowledge, our laboratory is the first to investigate the practicality of lip and eyelid movement because of the relatively small forces required, ease of implantation, ability to create animal models, and potential to significantly improve current methods of dynamic restoration of facial movement.8,12,13

We have been developing an artificial muscle–powered eyelid device to recreate blinking for adults and children with facial paralysis12 (Figure 1). We developed the eyelid blink and smile sling concepts by using an amalgam of current surgical procedures and philosophies to create a dynamic, controllable, and permanent alternative to more-invasive surgical options. The current reconstructive paradigm could benefit from the addition of artificial muscle because there are several shortcomings related to facial reanimation around the eye. For one, the gold or platinum eyelid-loading procedure results in a dyssynchronous and slow eyelid blink that relies on gravity to pull the eyelid closed. This can affect cosmesis but can also prevent the instinctive blink reflex to protect the cornea. A second shortcoming is that facial paralysis affects lacrimal function in several ways, including lacrimal oversecretion, eyelid ectropion, medial punctal eversion, and absence of a normal tear film.14 Finally, the creation of eyelid blinking with free tissue transfer has not shown great success.15 Unfortunately, current reconstructive methods fail to address all of these issues.

Place holder to copy figure label and caption
Graphic Jump Location

Figure 1. Conceptual illustration demonstrating a possible artificial muscle and power source design. A, Illustration of the potential applicability of a novel eyelid sling procedure using electroactive polymer artificial muscle (EPAM) to recreate eyelid blinking. The left upper eyelid sling is attached to the EPAM device, powered by a battery, and signaled by a myoelectric sensor (green wire) to the normally functioning orbicularis oculi muscle to coordinate eyelid blinking. B, An electroactive polymer device (*) includes an actuator (†) having an electroactive polymer (‡) in a biocompatible sheath (§). The electroactive polymer is attached the eyelid sling (∥) and energized by a power supply (¶). A myoelectric sensor (#) is in electrical contact with the actuator to provide a signal from the contralateral orbicularis oculi muscle to coordinate eyelid blinking. Reprinted with permission from Senders et al.12

To our knowledge, there have been no studies examining implantation of artificial muscle in a biological model. We seek to further develop a permanently implantable medical device that can create facial movement, such as a smile or opening and closing the eyelids. The device would ideally be interfaced to contralateral normal facial movement—allowing voluntary and symmetric eyelid control—and have a battery that is externally rechargeable without wires or connectors. This module would contain the artificial muscle actuator that would pull on a mechanical sling. We theorize that the EPAM device could be synchronized to the normal side of the face, but this has not been demonstrated. A myoelectric sensor threaded over the normal side could detect the voluntary facial impulse or involuntary blink and fire the artificial muscle at the same time12,13 A previous study13 showed that a sling device coupled to an EPAM actuator could be implanted in the upper and lower eyelids to create eyelid closure when the sling was activated.

The objectives of the current study were 2-fold: (1) to determine the durability and biocompatibility of the implanted EPAM device in a gerbil model and (2) to determine the degree of inflammation and fibrosis at the host tissue–artificial muscle interface. The gerbil was chosen because of its tolerance for implantation and our laboratory's experience with successful application of the head mount to safely direct the wires from the external power source to the EPAM device. A wireless, battery-operated system does not exist.1619 The general hypothesis is that an artificial muscle actuator has the potential to be an effective solution for restoring motion to nonfunctioning facial movement in children and adults with permanent facial paralysis. The estimated number of eyelid blinks that the potential EPAM eyelid sling would need to generate can be calculated by considering a normal 6 to 15 blinks/min20 for 16 waking hours a day for 365 days, equaling up to 5 million cycles of motion per year.

DEVICES AND POWER SUPPLY

Gerbils received implanted EPAM devices that were stimulated by an external power supply, which allowed for adjustment and monitoring of electrical input. The connection between the power supply and each gerbil was devised using experience by our laboratory in the gerbil model in other applications.17,18 In brief, the technique for external power connection to the gerbil involved screw fixation of a 3-mm spring to the posterior skull, bone cement to protect the attachment, and an electroconductive swivel connection (commutator) between the spring and the top of the cage. This allowed the animal to move about the cage without applying torque to the EPAM device (Figure 2). A function generator created a square wave input into the EPAM device, generating movement of the implant that could be palpated under the skin of the gerbil.

Place holder to copy figure label and caption
Graphic Jump Location

Figure 2. Experimental setup. A, Photograph of gerbil with head mount in place with wires within the spring extending from the animal's head. B, Design of the commutator (electroconductive swivel) unit allowing connection of the electroactive polymer artificial muscle (EPAM) device to the signal generator. C, Signal generator setup for stimulation of the EPAM units.

SURGICAL IMPLANTATION PROCEDURE FOR THE GERBIL MODEL

The surgical procedure for this experiment was derived from previous studies1619 on surgical implantation in gerbils. In brief, the animals were anesthetized with a mixture of ketamine hydrochloride, 80 mg/kg, and xylazine hydrochloride, 20 mg/kg. The fur over the skull was removed from the anterior limit of the eyes to the middle of the neck. The area was disinfected with povidone-iodine solution followed by alcohol, 70%; this was repeated 3 times. A parasagittal scalp incision was created, exposing the skull. A 1.0-cm-diameter section of periosteum was removed posterior to the coronal suture. A 1.2-mm-diameter burr hole was made lateral to the sagittal sinus for placement of a stainless steel anchoring screw (size 0-80, 0.32 cm). The free ends of the insulated EPAM stimulation wire were passed through a 3-mm-diameter stainless steel spring (to prevent chewing by the gerbil), one end of which was attached to the screw. Dental cement was then applied to the cranium to cover the stimulation wire and screw spring. The free end of the spring was connected to the electroconductive swivel (commutators) at the top of the cage to afford the gerbil free range of motion. These wires were connected to the signal function generator (271-U; Fluke Inc) and external high-voltage power supply amplifier controller (Model 610E; Trek, Inc).

The implanted EPAM devices were activated on initial implantation until euthanasia at 1 KV, 60 contractions per minute (1 Hz) for 24 h/d. The function of the EPAM device was monitored with the function generator wave output and by direct palpation of the device. Signal output data were recorded for 40 days after implantation. The schedule for euthanasia was based on when the EPAM device signal failed until day 40. When implants failed, they were removed from the signal generator, the animal was euthanized, and tissue was harvested for histologic analysis. Animals in which the units continued to function at the end of the 40 days were removed from the signal generator and euthanized between 40 and 65 days after implantation. Euthanasia was performed by injection of pentobarbital sodium, 200 mg/kg, and postmortem harvesting of the device and all surrounding tissue was completed to allow histologic assessment. The duration of activity of the actuators was analyzed with a Kaplan-Meier survival analysis. Statistical analysis was completed with commercial software (SPSS version 13; SPSS Inc).

HISTOLOGIC ANALYSIS

Specimens were fixed with paraformaldehyde, 4%, in buffered saline. After fixation, tissue was embedded in plastic, sectioned at 1 μm, placed on slides, and stained with toluidine blue and basic fuchsin stains. Sections were scored on the degree of acute and chronic inflammation, on the degree of fibrosis, and for the presence of a stromal reaction. Each histologic factor was graded on a 0 (absent) to 4 (severe) scale, as previously described.21

NUMBER OF ANIMALS

There were no data to support a power analysis in this pilot study. Animal numbers were derived on the basis of similar studies17,18 conducted on gerbils by our laboratory. All studies were completed under protocols approved by the Institutional Animal Care Use and Advisory Committee (07-13014) at the University of California, Davis.

Thirteen animals received implanted devices. There were no perioperative deaths and all gerbils recovered without problems. Twelve animals survived without difficulty until euthanasia between 19 and 65 days after implantation. One animal died 23 days after implantation resulting from a hemorrhage that occurred when the head mount screw fixation dislodged. No animals had overt evidence of infection.

Implanted devices functioned continuously for the 40 days of the experiment (mean, 30.3 days; range, 19-40 days after implantation) (Figure 3). Units were stimulated continuously by the signal generator at 60 cycles per minute throughout the 40-day study. Units sustained a mean of 2.6 × 106 cycles during the study (range, 1.6 × 106 to 3.5 × 106 cycles).

Place holder to copy figure label and caption
Graphic Jump Location

Figure 3. Duration of electroactive polymer artificial muscle (EPAM) device function. Device failure was attributable to electrical wire connections between head mount and commutator unit in all cases. Kaplan-Meier curve illustrating time to EPAM device motion detection failure. Mean duration was 30.3 days.

Histologic preparations of the implants examining the tissue-implant interface demonstrated the expected finding that, in the early phases of healing after implantation, there was a significant acute inflammatory infiltrate that receded after initial healing had completed (Figure 4). There was no evidence of prolonged polymorphonuclear cell inflammatory cell infiltration within the fibrous capsule surrounding the infiltrate by 40 days of implant life within the gerbil and remained absent up to 65 days after implantation. Chronic inflammation was minimal and a mild fibrous capsule formed around the implant. The units were also evaluated for evidence of breakdown of the silicone coating or infiltration of tissue into the silicone; none was observed.

Place holder to copy figure label and caption
Graphic Jump Location

Figure 4. Histologic analysis of the implant-tissue interface. A, Photograph of a section through the center of an in situ embedded implant. Arrows mark the upper surface of the skin and the lower boundary of the developing capsule. 1, Electroactive polymer artificial muscle (EPAM). 2, The upper and lower wire leads provide current to polymer. 3, The inner and outer layers of the silicone covering. The distance between each black line at the bottom of the figure is 1 mm. B and C, Representative histologic cross-sections of the implant-tissue interface with toluidine blue and basic fuchsin stains at original magnification ×4 and ×40, respectively. The asterisk represents the section of image B from which image C was obtained. D, Graph representing the grade of inflammatory response within the tissue surrounding the EPAM implants at days 23 through 65 after implantation. Two animals were maintained to day 65 to evaluate tissue-implant response. D indicates dorsal skin surface; V, ventral skin surface.

This study represents the first evaluation of the biocompatibility and durability of EPAM devices in an animal model. The results showed that implanted EPAM devices were safe, without stimulating a significant immune response or predisposing animals to infection. The implanted EPAM devices were stable within the subcutaneous site, and no obvious violation of the integrity of the devices was seen. The devices could be continuously stimulated without adverse effects on the animal for up to 40 days. The EPAM devices functioned for as many as 3.5 × 106 cycles, which would be equivalent to providing more than a year of continuous eyelid blinking for a patient with facial paralysis.

The next phase of experiments will aim to measure (using tensiometry and videography) the strength, durability, and reproducibility of movement created in vivo by artificial muscle. We seek to understand how the fibrotic wound healing process and fibrous capsule, which we observed in this preliminary experiment, will affect the functional movement of the eyelid sling mechanism and EPAM device function during a longer study.

Patients with facial paralysis experience significant functional and psychological impairment. Currently, several nerve grafting and muscle transferring surgeries are available; however, limitations in the quality of the results and the morbidity of the procedures demand that a better option be established. Artificial muscle to recreate facial movement is an exciting new concept that has drawn attention from multiple research forums, national news sources, and even family support groups. This study begins the process of understanding the long-term interactions between host tissue and artificial muscle. By ensuring the safety of these implanted devices, we hope to move forward in determining the functional abilities of these implants.

We had 1 animal death due to a hemorrhage after the sagittal sinus was exposed when the head mount detached from the skull on day 23. There did not appear to be any evidence of infection in this animal at necropsy. None of the other animals developed evidence of infection at the site of implantation, despite the wires exiting the skin from the implant.

This study has some important limitations. The EPAM devices, as designed for this study, did not impart any measurable actuation of anatomic structures. The life span of these units when increased stroke and force are created needs to be measured. We do not believe that this limitation affects the interpretation of the findings in light of the project's aims. Previous work22 with EPAM showed no detriment of effect with repeated stimulation while imparting force on a load ex vivo. This short-term implantation period may not reflect the fibrotic reaction to the EPAM device and possible sling material. Determining the functional capabilities of the implanted EPAM devices with regard to the life span while performing their expected functions will be accomplished in a study in a larger mammal (Yucatan pig) model being developed to test the eyelid sling mechanism. These studies will also shed light on the effects of fibrosis and scarring that will envelop the implanted devices and evaluate how detrimental these processes are on their function. Additionally, a self-contained battery-operated EPAM device has not yet been created.

The technology for long-lasting, implantable artificial muscle devices in humans is still in its infancy and needs considerable advancements in engineering before such devices are readily available. The eyelid sling, on which our laboratory has previously published data,12 provides an avenue for the use of such implantable devices in humans, although this represents only one of many potential applications. Studies are under way to establish an entirely implantable eyelid sling in a cadaver model. Before EPAM devices can be developed for use in humans, several obstacles must be overcome, including the need for smaller, more efficient EPAM device specifications, an internal power source, electromyographic sensors, and the interface between the EPAM device and sling mechanisms. Such devices would require implanted batteries with long life spans. Recent work23 in the nanotechnology field could make this a reality in the near future. Other recent work2427 suggests that electromyographic signals can be used to control devices, such as intelligent wheelchairs and advanced prostheses. In addition, recent studies28 have shown the feasibility of stimulation for a paralyzed face after detection of electromyographic signals from the normal side of the face.

In conclusion, we have shown in this study that implanted artificial muscle devices can function to create motion. The devices did not stimulate a significant inflammatory response or produce adverse effects. Further studies will be developed to explore the applicability of EPAM devices for restoration of facial motions in the population with irreversible facial paralysis. We believe this will be a safe and viable option, but further engineering designs and adaptation for use in humans remain obstacles to be overcome.

Correspondence: Travis T. Tollefson, MD, MPH, Department of Otolaryngology–Head and Neck Surgery, University of California, Davis Medical Center, 2521 Stockton Blvd, Ste 7200, Sacramento, CA 95817 (tttollefson@gmail.com).

Accepted for Publication: May 2, 2012.

Published Online: September 17, 2012. doi:10.1001/archfacial.2012.696

Author Contributions:Study concept and design: Ledgerwood, Tinling, Senders, Wong-Foy, Prahlad, and Tollefson. Acquisition of data: Ledgerwood, Tinling, Wong-Foy, Prahlad, and Tollefson. Analysis and interpretation of data: Ledgerwood, Senders, Wong-Foy, Prahlad, and Tollefson. Drafting of the manuscript: Ledgerwood, Senders, and Tollefson. Critical revision of the manuscript for important intellectual content: Ledgerwood, Tinling, Senders, Wong-Foy, Prahlad, and Tollefson. Statistical analysis: Ledgerwood and Tollefson. Obtained funding: Senders and Tollefson. Administrative, technical, and material support: Ledgerwood, Tinling, Senders, Wong-Foy, Prahlad, and Tollefson. Study supervision: Tinling, Senders, and Tollefson.

Financial Disclosure: None reported.

Funding/Support: This work was supported in part by grant CMNTT09 from the Children's Miracle Network.

Previous Presentation: This work was presented at the annual meeting of the American Academy of Facial Plastic and Reconstructive Surgery; September 10, 2011; San Francisco, CA.

Online-Only Material: Listen to an author interview about this article, and others, at http://bit.ly/MW0LW4.

Jackson CG, von Doersten PG. The facial nerve: current trends in diagnosis, treatment, and rehabilitation.  Med Clin North Am. 1999;83(1):179-195, x
PubMed   |  Link to Article
Berghaus A, Neumann K, Schrom T. The platinum chain: a new upper-lid implant for facial palsy.  Arch Facial Plast Surg. 2003;5(2):166-170
PubMed   |  Link to Article
Hadlock TA, Malo JS, Cheney ML, Henstrom DK. Free gracilis transfer for smile in children: the Massachusetts Eye and Ear Infirmary experience in excursion and quality-of-life changes.  Arch Facial Plast Surg. 2011;13(3):190-194
PubMed   |  Link to Article
Harii K, Ohmori K, Torii S. Free gracilis muscle transplantation, with microneurovascular anastomoses for the treatment of facial paralysis: a preliminary report.  Plast Reconstr Surg. 1976;57(2):133-143
PubMed   |  Link to Article
Bae YC, Zuker RM, Manktelow RT, Wade S. A comparison of commissure excursion following gracilis muscle transplantation for facial paralysis using a cross-face nerve graft versus the motor nerve to the masseter nerve.  Plast Reconstr Surg. 2006;117(7):2407-2413
PubMed   |  Link to Article
Chuang DC, Mardini S, Lin SH, Chen HC. Free proximal gracilis muscle and its skin paddle compound flap transplantation for complex facial paralysis.  Plast Reconstr Surg. 2004;113(1):126-135
PubMed   |  Link to Article
Frey M, Giovanoli P, Tzou CH, Kropf N, Friedl S. Dynamic reconstruction of eye closure by muscle transposition or functional muscle transplantation in facial palsy.  Plast Reconstr Surg. 2004;114(4):865-875
PubMed   |  Link to Article
Terzis JK, Noah ME. Analysis of 100 cases of free-muscle transplantation for facial paralysis.  Plast Reconstr Surg. 1997;99(7):1905-1921
PubMed   |  Link to Article
Zuker RM, Goldberg CS, Manktelow RT. Facial animation in children with Möbius syndrome after segmental gracilis muscle transplant.  Plast Reconstr Surg. 2000;106(1):1-9
PubMed   |  Link to Article
Ashley S. Artificial muscles.  Sci Am. 2003;289(4):52-59
PubMed   |  Link to Article
Kornbluh RRP, Pei Q, Oh S, Joseph J. Ultrahigh strain response of field-actuated elastomeric polymers.  Proc SPIE. 2000;3987:51-64
Senders CW, Tollefson TT, Curtiss S, Wong-Foy A, Prahlad H. Force requirements for artificial muscle to create an eyelid blink with eyelid sling.  Arch Facial Plast Surg. 2010;12(1):30-36
PubMed   |  Link to Article
Tollefson TT, Senders CW. Restoration of eyelid closure in facial paralysis using artificial muscle: preliminary cadaveric analysis.  Laryngoscope. 2007;117(11):1907-1911
PubMed   |  Link to Article
Miller MT, Strömland K, Ventura L. Congenital aberrant tearing: a re-look.  Trans Am Ophthalmol Soc. 2008;106:100-116
PubMed
Angelos PC, Downs BW. Options for the management of forehead and scalp defects.  Facial Plast Surg Clin North Am. 2009;17(3):379-393
PubMed   |  Link to Article
Chole RA, Tinling SP, Leverentz E, McGinn MD. Inhibition of nitric oxide synthase blocks osteoclastic bone resorption in adaptive bone modeling.  Acta Otolaryngol. 1998;118(5):705-711
PubMed   |  Link to Article
Wolfman DE, Chole RA. Osteoclast stimulation by positive middle-ear air pressure.  Arch Otolaryngol Head Neck Surg. 1986;112(10):1037-1042
PubMed   |  Link to Article
Wolfman DE, Tinling SP. Systemic osteoclast activation by localized pressure.  Arch Otolaryngol Head Neck Surg. 1993;119(2):165-168
PubMed   |  Link to Article
Yeung AH, Tinling SP, Brodie HA. Inhibition of post-meningitic cochlear injury with cerebrospinal fluid irrigation.  Otolaryngol Head Neck Surg. 2006;134(2):214-224
PubMed   |  Link to Article
Doughty MJ. Further assessment of gender- and blink pattern-related differences in the spontaneous eyeblink activity in primary gaze in young adult humans.  Optom Vis Sci. 2002;79(7):439-447
PubMed   |  Link to Article
Macleod TM, Williams G, Sanders R, Green CJ. Histological evaluation of Permacol as a subcutaneous implant over a 20-week period in the rat model.  Br J Plast Surg. 2005;58(4):518-532
PubMed   |  Link to Article
Pelrine R, Kornbluh R, Pei Q, Joseph J. High-speed electrically actuated elastomers with strain greater than 100%.  Science. 2000;287(5454):836-839
PubMed   |  Link to Article
Reddy AL, Shaijumon MM, Gowda SR, Ajayan PM. Coaxial MnO2/carbon nanotube array electrodes for high-performance lithium batteries.  Nano Lett. 2009;9(3):1002-1006
PubMed   |  Link to Article
Ju JS, Shin Y, Kim EY. Vision based interface system for hands free control of an intelligent wheelchair.  J Neuroeng Rehabil. 2009;6:33Link to Article
PubMed   |  Link to Article
Choi K. Control of a vehicle with EEG signals in real-time and system evaluation.  Eur J Appl Physiol. 2012;112(2):755-766
PubMed   |  Link to Article
Merrill DR, Lockhart J, Troyk PR, Weir RF, Hankin DL. Development of an implantable myoelectric sensor for advanced prosthesis control.  Artif Organs. 2011;35(3):249-252
PubMed   |  Link to Article
Frigerio A, Cavallari P. A closed-loop stimulation system supplemented with motoneurone dynamic sensitivity replicates natural eye blinks.  Otolaryngol Head Neck Surg. 2012;146(2):230-233
PubMed   |  Link to Article
Kurita M, Takushima A, Muraoka Y, Shiraishi T, Harii K. Feasibility of bionic reanimation of a paralyzed face: a preliminary study of functional electrical stimulation of a paralyzed facial muscle controlled with the electromyography of the contralateral health hemiface.  Plast Reconstr Surg. 2010;126:81e-83e
PubMed   |  Link to Article

Figures

Place holder to copy figure label and caption
Graphic Jump Location

Figure 1. Conceptual illustration demonstrating a possible artificial muscle and power source design. A, Illustration of the potential applicability of a novel eyelid sling procedure using electroactive polymer artificial muscle (EPAM) to recreate eyelid blinking. The left upper eyelid sling is attached to the EPAM device, powered by a battery, and signaled by a myoelectric sensor (green wire) to the normally functioning orbicularis oculi muscle to coordinate eyelid blinking. B, An electroactive polymer device (*) includes an actuator (†) having an electroactive polymer (‡) in a biocompatible sheath (§). The electroactive polymer is attached the eyelid sling (∥) and energized by a power supply (¶). A myoelectric sensor (#) is in electrical contact with the actuator to provide a signal from the contralateral orbicularis oculi muscle to coordinate eyelid blinking. Reprinted with permission from Senders et al.12

Place holder to copy figure label and caption
Graphic Jump Location

Figure 2. Experimental setup. A, Photograph of gerbil with head mount in place with wires within the spring extending from the animal's head. B, Design of the commutator (electroconductive swivel) unit allowing connection of the electroactive polymer artificial muscle (EPAM) device to the signal generator. C, Signal generator setup for stimulation of the EPAM units.

Place holder to copy figure label and caption
Graphic Jump Location

Figure 3. Duration of electroactive polymer artificial muscle (EPAM) device function. Device failure was attributable to electrical wire connections between head mount and commutator unit in all cases. Kaplan-Meier curve illustrating time to EPAM device motion detection failure. Mean duration was 30.3 days.

Place holder to copy figure label and caption
Graphic Jump Location

Figure 4. Histologic analysis of the implant-tissue interface. A, Photograph of a section through the center of an in situ embedded implant. Arrows mark the upper surface of the skin and the lower boundary of the developing capsule. 1, Electroactive polymer artificial muscle (EPAM). 2, The upper and lower wire leads provide current to polymer. 3, The inner and outer layers of the silicone covering. The distance between each black line at the bottom of the figure is 1 mm. B and C, Representative histologic cross-sections of the implant-tissue interface with toluidine blue and basic fuchsin stains at original magnification ×4 and ×40, respectively. The asterisk represents the section of image B from which image C was obtained. D, Graph representing the grade of inflammatory response within the tissue surrounding the EPAM implants at days 23 through 65 after implantation. Two animals were maintained to day 65 to evaluate tissue-implant response. D indicates dorsal skin surface; V, ventral skin surface.

Tables

References

Jackson CG, von Doersten PG. The facial nerve: current trends in diagnosis, treatment, and rehabilitation.  Med Clin North Am. 1999;83(1):179-195, x
PubMed   |  Link to Article
Berghaus A, Neumann K, Schrom T. The platinum chain: a new upper-lid implant for facial palsy.  Arch Facial Plast Surg. 2003;5(2):166-170
PubMed   |  Link to Article
Hadlock TA, Malo JS, Cheney ML, Henstrom DK. Free gracilis transfer for smile in children: the Massachusetts Eye and Ear Infirmary experience in excursion and quality-of-life changes.  Arch Facial Plast Surg. 2011;13(3):190-194
PubMed   |  Link to Article
Harii K, Ohmori K, Torii S. Free gracilis muscle transplantation, with microneurovascular anastomoses for the treatment of facial paralysis: a preliminary report.  Plast Reconstr Surg. 1976;57(2):133-143
PubMed   |  Link to Article
Bae YC, Zuker RM, Manktelow RT, Wade S. A comparison of commissure excursion following gracilis muscle transplantation for facial paralysis using a cross-face nerve graft versus the motor nerve to the masseter nerve.  Plast Reconstr Surg. 2006;117(7):2407-2413
PubMed   |  Link to Article
Chuang DC, Mardini S, Lin SH, Chen HC. Free proximal gracilis muscle and its skin paddle compound flap transplantation for complex facial paralysis.  Plast Reconstr Surg. 2004;113(1):126-135
PubMed   |  Link to Article
Frey M, Giovanoli P, Tzou CH, Kropf N, Friedl S. Dynamic reconstruction of eye closure by muscle transposition or functional muscle transplantation in facial palsy.  Plast Reconstr Surg. 2004;114(4):865-875
PubMed   |  Link to Article
Terzis JK, Noah ME. Analysis of 100 cases of free-muscle transplantation for facial paralysis.  Plast Reconstr Surg. 1997;99(7):1905-1921
PubMed   |  Link to Article
Zuker RM, Goldberg CS, Manktelow RT. Facial animation in children with Möbius syndrome after segmental gracilis muscle transplant.  Plast Reconstr Surg. 2000;106(1):1-9
PubMed   |  Link to Article
Ashley S. Artificial muscles.  Sci Am. 2003;289(4):52-59
PubMed   |  Link to Article
Kornbluh RRP, Pei Q, Oh S, Joseph J. Ultrahigh strain response of field-actuated elastomeric polymers.  Proc SPIE. 2000;3987:51-64
Senders CW, Tollefson TT, Curtiss S, Wong-Foy A, Prahlad H. Force requirements for artificial muscle to create an eyelid blink with eyelid sling.  Arch Facial Plast Surg. 2010;12(1):30-36
PubMed   |  Link to Article
Tollefson TT, Senders CW. Restoration of eyelid closure in facial paralysis using artificial muscle: preliminary cadaveric analysis.  Laryngoscope. 2007;117(11):1907-1911
PubMed   |  Link to Article
Miller MT, Strömland K, Ventura L. Congenital aberrant tearing: a re-look.  Trans Am Ophthalmol Soc. 2008;106:100-116
PubMed
Angelos PC, Downs BW. Options for the management of forehead and scalp defects.  Facial Plast Surg Clin North Am. 2009;17(3):379-393
PubMed   |  Link to Article
Chole RA, Tinling SP, Leverentz E, McGinn MD. Inhibition of nitric oxide synthase blocks osteoclastic bone resorption in adaptive bone modeling.  Acta Otolaryngol. 1998;118(5):705-711
PubMed   |  Link to Article
Wolfman DE, Chole RA. Osteoclast stimulation by positive middle-ear air pressure.  Arch Otolaryngol Head Neck Surg. 1986;112(10):1037-1042
PubMed   |  Link to Article
Wolfman DE, Tinling SP. Systemic osteoclast activation by localized pressure.  Arch Otolaryngol Head Neck Surg. 1993;119(2):165-168
PubMed   |  Link to Article
Yeung AH, Tinling SP, Brodie HA. Inhibition of post-meningitic cochlear injury with cerebrospinal fluid irrigation.  Otolaryngol Head Neck Surg. 2006;134(2):214-224
PubMed   |  Link to Article
Doughty MJ. Further assessment of gender- and blink pattern-related differences in the spontaneous eyeblink activity in primary gaze in young adult humans.  Optom Vis Sci. 2002;79(7):439-447
PubMed   |  Link to Article
Macleod TM, Williams G, Sanders R, Green CJ. Histological evaluation of Permacol as a subcutaneous implant over a 20-week period in the rat model.  Br J Plast Surg. 2005;58(4):518-532
PubMed   |  Link to Article
Pelrine R, Kornbluh R, Pei Q, Joseph J. High-speed electrically actuated elastomers with strain greater than 100%.  Science. 2000;287(5454):836-839
PubMed   |  Link to Article
Reddy AL, Shaijumon MM, Gowda SR, Ajayan PM. Coaxial MnO2/carbon nanotube array electrodes for high-performance lithium batteries.  Nano Lett. 2009;9(3):1002-1006
PubMed   |  Link to Article
Ju JS, Shin Y, Kim EY. Vision based interface system for hands free control of an intelligent wheelchair.  J Neuroeng Rehabil. 2009;6:33Link to Article
PubMed   |  Link to Article
Choi K. Control of a vehicle with EEG signals in real-time and system evaluation.  Eur J Appl Physiol. 2012;112(2):755-766
PubMed   |  Link to Article
Merrill DR, Lockhart J, Troyk PR, Weir RF, Hankin DL. Development of an implantable myoelectric sensor for advanced prosthesis control.  Artif Organs. 2011;35(3):249-252
PubMed   |  Link to Article
Frigerio A, Cavallari P. A closed-loop stimulation system supplemented with motoneurone dynamic sensitivity replicates natural eye blinks.  Otolaryngol Head Neck Surg. 2012;146(2):230-233
PubMed   |  Link to Article
Kurita M, Takushima A, Muraoka Y, Shiraishi T, Harii K. Feasibility of bionic reanimation of a paralyzed face: a preliminary study of functional electrical stimulation of a paralyzed facial muscle controlled with the electromyography of the contralateral health hemiface.  Plast Reconstr Surg. 2010;126:81e-83e
PubMed   |  Link to Article

Correspondence

CME
Also Meets CME requirements for:
Browse CME for all U.S. States
Accreditation Information
The American Medical Association is accredited by the Accreditation Council for Continuing Medical Education to provide continuing medical education for physicians. The AMA designates this journal-based CME activity for a maximum of 1 AMA PRA Category 1 CreditTM per course. Physicians should claim only the credit commensurate with the extent of their participation in the activity. Physicians who complete the CME course and score at least 80% correct on the quiz are eligible for AMA PRA Category 1 CreditTM.
Note: You must get at least of the answers correct to pass this quiz.
Please click the checkbox indicating that you have read the full article in order to submit your answers.
Your answers have been saved for later.
You have not filled in all the answers to complete this quiz
The following questions were not answered:
Sorry, you have unsuccessfully completed this CME quiz with a score of
The following questions were not answered correctly:
Commitment to Change (optional):
Indicate what change(s) you will implement in your practice, if any, based on this CME course.
Your quiz results:
The filled radio buttons indicate your responses. The preferred responses are highlighted
For CME Course: A Proposed Model for Initial Assessment and Management of Acute Heart Failure Syndromes
Indicate what changes(s) you will implement in your practice, if any, based on this CME course.

Multimedia

Some tools below are only available to our subscribers or users with an online account.

2,091 Views
3 Citations
×

Related Content

Customize your page view by dragging & repositioning the boxes below.

Related Multimedia

Author Interview

audio player

Articles Related By Topic
Related Collections
Jobs