Original Article |

Long-term Efficacy of Biomodeled Polymethyl Methacrylate Implants for Orbitofacial Defects FREE

Michael J. Groth, MD; Aparna Bhatnagar, MD; William J. Clearihue, PhD; Robert A. Goldberg, MD; Raymond S. Douglas, MD, PhD
[+] Author Affiliations

Correspondence: Raymond S. Douglas, MD, PhD, Orbital and Ophthalmic Plastic Surgery Division, Jules Stein Eye Institute, 100 Stein Plaza, Room 2-267, Los Angeles, CA 90095 (raymonddouglasmd@yahoo.com).


Author Affiliations: Orbit and Ophthalmic Plastic Surgery Division, Jules Stein Eye Institute, Los Angeles (Drs Groth, Bhatnagar, Goldberg, and Douglas), Brighton Laser and Surgery Institute, Beverly Hills (Dr Groth), and Orbital and Ophthalmic Plastic and Reconstructive Surgery, Greater Los Angeles Veterans Administration Hospital, West Los Angeles (Drs Clearihue and Douglas), Calif.


Arch Facial Plast Surg. 2006;8(6):381-389. doi:10.1001/archfaci.8.6.381.
Text Size: A A A
Published online

Objective  To report the long-term efficacy of custom polymethyl methacrylate implants using high-resolution computed tomographic modeling in the reconstruction of complex orbitofacial defects secondary to trauma.

Methods  Nine patients with complex orbitofacial bone defects after trauma were evaluated for this retrospective, nonrandomized, noncomparative study. All the patients underwent reconstruction using custom, heat-cured polymethyl methacrylate implants. Patients were followed up postoperatively and evaluated for complications.

Results  Nine consecutive patients (5 men and 4 women) aged 28 to 63 years who underwent surgical reconstruction using prefabricated, heat-cured polymethyl methacrylate implants were included in the study. The interval between injury and presentation ranged from 1 month to 40 years. There were no significant complications, including infection, extrusion, or displacement of the implant. In all of the patients, wound healing was uneventful, with antibiotic drugs administered perioperatively. Mean follow-up was 4.3 years from the first visit (range, 6 months to 10 years).

Conclusions  Computed tomographic biomodeled, prefabricated, heat-cured polymethyl methacrylate implants are well tolerated in the long term. Their advantages include customized design, long-term biocompatibility, and excellent aesthetic results.

Figures in this Article

The accurate replication of complex human anatomy has long been sought for patients with severe orbitofacial defects. These abnormalities could result from trauma, congenital defects, or iatrogenic defects such as tumor removal. The main objective of reconstruction of orbitofacial defects is to restore anatomical integrity, providing proper eyelid and facial function and cosmetic improvement. Secondary reconstructions pose a particular challenge to the surgeon. Severe trauma and the resultant multicontoured disruption of tissues causes extensive scarring, loss of anatomical landmarks, and loss of functional tissue. Secondary reconstruction can exacerbate all of these conditions.

Many autologous and allogenic implant materials have been used with varying success and indications. In general, autologous and alloplastic materials, which are modeled intraoperatively, have failed to create ideal 3-dimensional (3-D) facial and orbital contours. In contrast, high-resolution preoperative modeling facilitates precise implant construction and placement, thus providing improved facial contour and implant stability. Concerns regarding alloplastic material placement include long-term implant stability and risk of infection.1 Herein we report the successful long-term efficacy of custom polymethyl methacrylate (PMMA) implants using high-resolution computed tomographic (CT) modeling in the reconstruction of complex orbitofacial defects secondary to trauma.

PATIENTS

Nine consecutive patients who underwent reconstruction of orbitofacial defects secondary to trauma using CT biomodeled implants were included in the study. Criteria for the use of a custom CT biomodel included severe bony trauma with multicontoured orbitofacial defects and unsatisfactory primary reconstruction. Preoperative considerations included the degree of trauma, previous surgery, tissue volume, and functional deficits. Measurements of the pupillary distance, degree of enophthalmos, displacement of zygoma, and orbital rim were included. In all of the cases, patients underwent 3-D CT of the orbits and construction of a custom PMMA implant to duplicate the anatomical defect.

HIGH-RESOLUTION CT BIOMODELED CUSTOM PMMA IMPLANTS

Acquisition by CT and transfer and evaluation of patient data were performed according to principles established in preliminary studies on geometrically standardized models and the production of implants for bony models.2 Generation of a physical model (BMI Biomedical Modeling Inc, Boston, Mass) from CT data required 4 basic steps. The first step was acquisition of CT data using specified protocols (Figure 1). The second step was delineation of a CT data subset, which described the model itself (Figure 2). The third step involved data translation into a rapid prototyping machine to form the 3-D model. Using these data, an accurate prefabricated PMMA implant was constructed using a 6-step process.2 For best results, strict protocols on patient movement and technical considerations were followed (a copy of the BMI Biomedical Modeling Inc protocol can be obtained via e-mail at http://www.biomodel.com). Actual scanning time was kept as short as possible, with 0.5- to 1.0-mm slices.

Place holder to copy figure label and caption
Figure 1.

Computed tomographic scan showing all 3 dimensions (sagittal, coronal, and axial) of the orbitofacial defect as a preliminary step in the formation of a biomodel.

Graphic Jump Location
Place holder to copy figure label and caption
Figure 2.

Photographs of a 3-dimensional (3-D) computed tomographic biomodel with a polymethyl methacrylate (PMMA) implant. A, Front view of the biomodel (shown in violet) with the PMMA implant (shown in green) covering the orbital defect. B, Diagonal view of the same biomodel and implant demonstrating the 3-D effect and showing depth and dimension of the implant, which is fitted into the defect. The prefabricated PMMA implant (front [a] and diagonal [b] view) is constructed using computer-generated data and the biomodel. The generation of the 3-D biomodel is crucial, as it determines the exact shape and dimension of the future prefabricated PMMA implant.

Graphic Jump Location
DATA TRANSLATION

The technician identified the 3-D region of data describing the model, and the slices were reconstructed into a 3-D object. The reconstruction also interpolated between the slices, filling in the “gaps” within each single slice and producing a smoother model.

IMPLANT FABRICATION TECHNIQUE

Generation of an accurate PMMA implant from the 3-D CT model required several steps. Briefly, a thin layer of dental die stone was painted on the bony tissue side of the wax. After the die stone hardened, the mold was separated and the cavity was thoroughly cleaned. The PMMA non–cross-linked acrylic monomer and polymer were processed and packed into the stone mold. The mold was subjected to a 3-stage water bath heat-curing unit for 15 hours. Finally, the newly custom-fabricated, accurate PMMA implant and the 3-D CT biomodel were sterilized using ethylene oxide gas followed by aeration for at least 72 hours to remove residual ethylene oxide.

SURGERY

All the procedures were performed by 2 of us (M.J.G. and R.A.G.). The goal of surgery was to restore anatomical function, orbital volume, and facial contour. During reconstruction of the orbitofacial defects, care was taken to ensure precise implant placement. It was important that the implant did not rest on any high points of bone because this could dramatically alter orbital volume or contour. Appearance of a thin meniscus of fluid between the bone implant interfaces was indicative of proper placement of the implant (Figure 3A), with an even distribution of fluid over the entire surface of the bone under the clear implant (Figure 3B). If any high points were noted from either the PMMA implant or the bone, these were trimmed down at the time of surgery. After surgery, patients were observed for any signs of complications, including orbital hemorrhage, severe pain, discharge, displacement, and extrusion of the implant.

Place holder to copy figure label and caption
Figure 3.

Intraoperative photographs showing accuate placement of the prefabricated polymethyl methacrylate (PMMA) implant. A, Formation of a thin meniscus of fluid between the bone implant interface is seen, which confirms proper placement of the PMMA implant over the anatomical defect. B, Clear distribution of fluid is seen over the bone under the transparent PMMA implant.

Graphic Jump Location

Nine consecutive patients with complex orbitofacial bone and soft tissue injuries secondary to trauma were evaluated for reconstructive surgery. Their ages ranged from 28 to 63 years (mean, 48.7 years). There were 5 men and 4 women. The interval between injury and presentation was 1 month to 40 years (Table 1), during which time all of the patients underwent surgical reconstruction elsewhere. Procedures before reconstruction with PMMA included the use of autologous bone graft (2 patients), porous polyethylene implants (2 patients), titanium plate implants (2 patients), and soft tissue reconstruction with hard-palate eyelid grafting and silicone injection (1 patient each) (Table 2).

Table Graphic Jump LocationTable 1. Characteristics of Patients Who Underwent Reconstructive Procedures Using Custom, Prefabricated PMMA Implants
Table Graphic Jump LocationTable 2. Surgical Data of Patients Who Underwent Reconstruction Using PMMA Implants*

All of the patients demonstrated signs and symptoms of severe orbital trauma and cicatrisation, including facial asymmetry (3 patients), orbital asymmetry including hypoglobus (6 patients), and diplopia (5 patients). Patient characteristics and details of their injuries are summarized in Table 1. Ocular and orbital examination revealed decreased visual acuity in 6 of 8 patients, hypoglobus in 6, restrictive strabismus in 5, lower eyelid ectropion in 5, and enophthalmos in 4. Upper eyelid ptosis was present in 3 patients, superior sulcus deformity in 2, and exposure keratopathy in 2. One patient had traumatic mydriasis, aphakia, retinopathy and optic neuropathy, and frontal nerve hypoesthesia (Figure 4). After reconstruction using custom PMMA implants, no surgical complications were reported. Two procedures required minor intraoperative modification, such as grinding off excessive PMMA implant or cutting high points of bone. All of the implants were fixed using screws that were counterfitted (below the level of the implant).This prevents any projection of the screws, thus eliminating palpation externally. The surgical details are included in Table 2.

Place holder to copy figure label and caption
Figure 4.

Clinical features of the 9 patients who underwent surgical reconstruction of traumatic orbitofacial defects using polymethyl methacrylate implants. “Others” includes traumatic mydriasis, aphakia, retinopathy with optic neuropathy, and hypoesthesia over the frontal nerve region. VA indicates visual acuity; SS, superior sulcus.

Graphic Jump Location

Figure 5 demonstrates signs of orbitofacial deformity and scarring secondary to a boating accident 3 years earlier. Orbital trauma included a right blowout fracture, a right tripod fracture, and lateral bowing of the medial wall of the maxillary antrum with the absence of its superior wall, in addition to multiple facial injuries. Previous surgical reconstruction included multiple eyelid operations and scleral buckling for retinal detachment. Preoperative assessment demonstrated 2 mm of right enophthalmos. Eyelid examination revealed 1 mm of right lower eyelid ectropion, with 2 mm of inferior displacement of the right lateral canthal angle, superior sulcus deformity, and cicatricial orbitopathy. The patient underwent high-resolution CT biomodeling and custom implant construction. Surgical reconstruction using a PMMA implant included right orbital reconstruction with repair of the orbital blowout fracture and orbital rim fracture. Additional procedures included right upper eyelid ptosis repair and right inferior conjunctivoplasty using a free hard-palate mucous membrane graft with right canthoplasty. Intraoperatively, constant comparison was made between the patient's anatomy and that of the model, which was placed on the table adjacent to the operating field. The implant fit well, with minor modification necessary along the posterior and medial aspects of the orbital floor. The motility of the eye was constantly checked during the procedure to prevent entrapment of the septae or extraocular muscle. The postoperative photograph demonstrates good improvement of the superior sulcus defect and improvement of enophthalmos and eyelid contour and function (Figure 6).

Place holder to copy figure label and caption
Figure 5.

Photograph demonstrating significant posttraumatic enophthalmos, superior sulcus deformity, and cicatricial orbitopathy. This patient underwent high-resolution computed tomographic biomodeling, construction of a prefabricated polymethyl methacrylate implant, and reconstructive surgery with implantation.

Graphic Jump Location
Place holder to copy figure label and caption
Figure 6.

Postoperative photograph showing significant correction of the superior sulcus deformity, hypoglobus, and resolving enophthalmos. The patient was followed up for 10 years and showed no signs of any complication.

Graphic Jump Location

Figure 7A shows a patient with severe facial and orbital trauma after a fall from a 50-ft height. Extensive orbital trauma included fractures of the right frontal bone, orbital floor, and zygomatic arch (posterior and lateral displacement consistent with trimalar fracture) and inferior displacement of orbital soft tissue. Preoperative assessment of the right eye revealed 3.5 mm of hypoglobus, with depression of malar eminence, superior sulcus deformity, andorbital volume deficit with 3 mm of enophthalmos. There was hypoesthesia of the ophthalmic division of the trigeminal cranial nerve. Motility examination showed limitation of upgaze and downgaze in the right eye. The patient underwent CT biomodeling and custom implant construction. The orbital anatomical reconstructions with implants are demonstrated in Figure 8A. Because the patient had a superior orbital rim defect and an inferior orbital floor defect, 2 implants were designed (Figure 8B and C). Figure 8D demonstrates typical lock-and-key fit to anatomical structures. The same precise fit was achieved intraoperatively. A superior eyelid crease incision and an inferior transcon junctival incision with subperiosteal dissection were used to achieve access to the bony defects (Figure 9 and Figure 10). Postoperative assessment demonstrates significant improvement of the hypoglobus, superior sulcus defect, and enophthalmos (Figure 7B and C).

Place holder to copy figure label and caption
Figure 7.

Photographs of a patient with severe facial and orbital trauma after a fall from a 50-ft height. A, Significant posttraumatic hypoglobus, restrictive strabismus, enophthalmos, and orbital volume defect of the right eye. The patient underwent multiple reconstructive procedures of the face and orbit but still experienced diplopia and facial asymmetry before undergoing reconstructive surgery with implantation of a prefabricated, custom polymethyl methacrylate implant. B, Early postoperative photograph showing minimal orbital inflammation, with sutures at the superior eyelid crease and lateral canthal angle after reconstructive surgery. C, At 1-year follow-up there is improvement of the hypoglobus, superior sulcus deformity, and enophthalmos.

Graphic Jump Location
Place holder to copy figure label and caption
Figure 8.

Photographs showing computed tomographic biomodeling of the patient's anatomical defects using polymethyl methacrylate (PMMA) implants placed over the defect. Two separate PMMA implants were constructed for superior and inferior bony defects (A). Prefabricated custom PMMA implants were designed for the superior orbital rim (B) and the orbital floor (C), with a lock-and-key fit to the anatomical structures (D).

Graphic Jump Location
Place holder to copy figure label and caption
Figure 9.

Intraoperative photographs showing placement of a custom polymethyl methacrylate implant for a superior orbital rim defect. A, A superior eyelid crease incision was used to expose the fracture site. B, Fixation of the polymethyl methacrylate implant with metallic screws.

Graphic Jump Location
Place holder to copy figure label and caption
Figure 10.

Intraoperative photographs of a patient who underwent reconstruction of an orbital floor defect after trauma. A, An inferior transconjunctival incision was made with subperiosteal dissection to achieve access to the bony defect. B, Placement of a custom polymethyl methacrylate implant in the orbital floor defect.

Graphic Jump Location

Figure 11 demonstrates preoperative and postoperative photographs of a patient who underwent PMMA implant reconstruction of multiple bony injuries and eyelid lacerations after a motor vehicle crash 1 year before presentation. On examination, he demonstrated a healed malar fracture with hypoglobus and enophthalmos of 2 mm each in addition to some degree of midface descent and hollowing of the right malar and submalar regions (Figure 11A). The CT scan demonstrated a comminuted fracture of the right maxilla involving the anterior ramus of the orbital floor (Figure 12). A CT biomodel was designed that clearly illustrated the areas of bony defect in the orbital floor (Figure 13A). Accordingly, a PMMA implant was constructed (Figure 13A inset) to fully conceal the fracture site (Figure 13B). The patient withstood the surgery well and showed marked improvement after surgery (Figure 11B).

Place holder to copy figure label and caption
Figure 11.

Preoperative and postoperative photographs of a patient who underwent reconstruction using a polymethyl methacrylate implant for multiple bony injuries and eyelid lacerations after a motor vehicle crash. A, Preoperative photograph showing right hypoglobus and enophthalmos in addition to midface descent and hollowing of the malar region. B, Postoperative photograph at 8-month follow-up showing improvement of the hypoglobus and facial symmetry.

Graphic Jump Location
Place holder to copy figure label and caption
Figure 12.

Coronal computed tomographic scan of the face demonstrating a comminuted fracture of the right maxillary anterior wall with hemorrhagic opacification of the right maxillary sinus. Also seen is the fracture of the right orbital floor involving the anterior ramus.

Graphic Jump Location
Place holder to copy figure label and caption
Figure 13.

Photographs of the computed tomographic biomodel and custom prefabricated implant for reconstruction of an orbitofacial defect. A, Computed tomographic biomodel prepared for the patient with an orbital floor fracture. The polymethyl methacrylate implant is shown in the inset. B, Photograph showing accurate placement of the polymethyl methacrylate implant over the orbital floor defect.

Graphic Jump Location

Long-term data demonstrate that placement of orbital implants achieves reliable stability and long-term improvement in function with minimal risk of extrusion or implant infection. None of the patients in this study developed significant complications, including infection, extrusion, or displacement of the implant. All of the patients had minimal postoperative pain, and none developed hematoma or seroma. No surgical drains were used. In all of the patients, wound healing was uneventful, with antibiotics given perioperatively. Mean follow-up after reconstruction was 4.3 years (range, 6 months to 10 years). All of the patients had a routine follow-up with complete ophthalmologic evaluation once a week for a month, followed by once every 2 months for 6 months and quarterly annual visits thereafter.

Many types of implant materials have been used for reconstructive orbitofacial surgery. Distinction between the use of autogenous bone grafts and alloplastic materials has largely been based on issues concerning biocompatibility and availability. Autogenous materials such as bone grafts provide many advantages, including excellent biocompatibility and low infection and extrusion rates.3 However, these materials have pertinent limitations, such as the potential for a prolonged operative procedure,4 limited supply, significant resorption, donor site morbidity, and minimal malleability and customization. Craniofacial implants have been successfully made from materials such as PMMA and silicone polymers for more than 45 years.5 Prefabrication of PMMA offers customization while using an established biocompatible material. In 2 of the present patients, the previous autologous bone grafts had to be removed owing to significant resorption or displacement. Customizing autografts intraoperatively is difficult, and it is technically impossible to achieve ideal precision for eyelid and facial symmetry. Small asymmetries of 1 to 3 mm are not only aesthetically noticeable but can compromise eyelid function. In the patients described herein, periorbital facial symmetry was improved, as was eyelid function. Use of CT biomodeling for alloplastic materials is convenient and provides precise and functional reconstruction.

Prefabricated computer-generated implants made from carbon fiber–reinforced polymer and prebent titanium plates have also been used for cranioplasty but are expensive. The ideal alloplastic material should be biocompatible, inert, lightweight, rigid, and inexpensive. Overall, PMMA is well suited to custom orbital implantation and generates minimal thermal, electrical, and magnetic conductivity.6

Previous literature has raised concerns that alloplastic implants are prone to infection and explantation in the long term. A study by Jordan et al7 reported that the incidence of early and delayed complications varies from 0.4% to 7.0% with alloplastic implants. A study8 of 37 orbital reconstructions using titanium implants reported complications such as undercorrection, symptomatically palpable implants, and orbital infection requiring explantation. In another study,4 a 7% complication rate included tight orbital fit, gaze limitation, and eventual development of enophthalmos. Infections usually develop as a sequela to floor and medial wall orbital fractures. The proximity and disruption of the paranasal sinuses seem too significant for infection.9 Bioactive glass has been used in the reconstruction of the orbital floor. The authors noted complications such as ectropion and infraorbital nerve hypoesthesia in 2 of 28 patients in their series.10 Orbital abscess, recurrent infection leading to implant removal, postoperative infraorbital anesthesia, and palpable screws were some of the complications encountered in other studies11-12 using alloplastic implants, such as porous polyethylene and silicone. In addition, 2 case reports described orbital fracture reconstruction using woven polyamide, followed by fistula formation from the maxillary antrum and ethmoid air cells. Both cases were cured by removal of the implant and surrounding tissue.7 In the present study, although the sinuses were involved in 6 of 8 cases, no such complication occurred. The incidence of perioperative infection has been remarkably low for porous hydroxyapatite.13 This resistance has been attributed to antibiotic-containing blood filling the pores of the implant, thus preventing normal and pathologic flora from establishing infection in the implant.13-14 However, cases with hydroxyapatite implants have been reported to develop clinically significant deep-tissue infection.15 Although the number of patients in this study is too small to address the issue of infection rate, all of these cases were secondary reconstructions. In addition, CT biomodeling and prefabrication of implants in the present study provided geometric standardization and gapless technical precision and guaranteed an exact fit. In the present study, all of the screws were counterfitted (buried below the bony level) so they could not be palpated. This reflects the safety and effectiveness of the PMMA and ensures its long-term stability without the risk of displacement leading to infection and removal. Polymethyl methacrylate was shown to be well tolerated without presenting biological adverse effects such as foreign body reactions.16

In the present study, all of the patients were first seen many years after the trauma and primary reconstruction (1 patient presented 40 years after the trauma) with facial asymmetry and deformity. Reconstruction with custom biomodeled implants offered a practical alternative in complex cases. It requires preoperative CT according to specific parameters and can be accomplished in almost all radiology centers. Model construction is straightforward, with reduced operative time. All of the patients demonstrated long-term sustained improvement of facial deformities.

We conclude that biomodeled PMMA implants are well tolerated in the long term. Although this study is relatively small, it is the first time long-term data have been presented regarding the applicability of biomodeled PMMA implants. We believe that the advantages of customization and long-term efficacy reported in this article far outweigh potential complications in complex orbital reconstruction, which requires precise implant customization to provide function improvement.

Correspondence: Raymond S. Douglas, MD, PhD, Orbital and Ophthalmic Plastic Surgery Division, Jules Stein Eye Institute, 100 Stein Plaza, Room 2-267, Los Angeles, CA 90095 (raymonddouglasmd@yahoo.com).

Accepted for Publication: July 26, 2006.

Financial Disclosure: None reported.

Previous Presentation: This study was presented as a poster at the World Ophthalmology Congress; February 21, 2006; São Paulo, Brazil.

Eufinger  HWehmoller  MHarders  AHeuser  L. Prostheses for the reconstruction of skull defects. Int J Oral Maxillofac Surg 1995;24104- 110
PubMed
Clearihue  WJGroth  MJNoel  JC. PMMA implant fabrication, utilizing the Photo-polymer Solid Ground Forming Process. J Facial Somato Prosthet 1998;4105- 118
Taub  PJRudkin  GHClearihue  WJ  IIIMiller  TA. Prefabricated alloplastic implants for cranial defects. Plast Reconstr Surg 2003;1111233- 1240
PubMed
Janecka  IP New reconstructive technologies in skull base surgery: role of titanium mesh and porous polyethylene. Arch Otolaryngol Head Neck Surg 2000;126396- 401
PubMed
Steinhauser  EWHardt  N Secondary reconstruction of cranial defects. J Maxillofac Surg 1977;5192- 198
PubMed
Chiarini  LFigurelli  SPollastri  G  et al.  Cranioplasty using acrylic material: a new technical procedure. J Craniomaxillofac Surg 2004;325- 9
PubMed
Jordan  DROnge  PSAnderson  RLPatrinely  JRNerad  JA. Complications associated with alloplastic implants used in orbital fracture repair. Ophthalmology 1992;991600- 1608
PubMed
Rubin  PABilyk  JRShore  JW. Orbital reconstruction using porous polyethylene sheets. Ophthalmology 1994;1011697- 1708
PubMed
McNab  AA Sino-orbital fistula: two case reports. Clin Experiment Ophthalmol 2000;28324- 325
PubMed
Kinnunen  IAitasalo  KPollonen  MVarpula  M. Reconstruction of orbital floor fractures using bioactive glass. J Craniomaxillofac Surg 2000;28229- 234
PubMed
Ng  SGMadill  SAInkster  CFMaloof  AJLeatherbarrow  B. Medpor porous polyethylene implants in orbital blowout fracture repair. Eye 2001;15578- 582
PubMed
Mauriello  JA  JrHargrave  SYee  SMostafavi  RKapila  R Infection after insertion of alloplastic orbital floor implants. Am J Ophthalmol 1994;117246- 252
PubMed
Rosen  HM Porous, block hydroxyapatite as an interpositional bone graft substitute in orthognathic surgery. Plast Reconstr Surg 1989;83985- 990
PubMed
Merritt  KShafer  JWBrown  S Implant site infection rates with porous and dense materials. J Biomed Mater Res 1979;13101- 108
PubMed
Rosen  HM The response of porous hydroxyapatite to contiguous tissue infection. Plast Reconstr Surg 1991;881076- 1080
PubMed
Luparello  DBruschi  SVerna  G  et al.  Cranioplasty with polymethylmethacrylate: the clinico-statistical considerations. Minerva Chir 1998;53575- 579
PubMed

Figures

Place holder to copy figure label and caption
Figure 1.

Computed tomographic scan showing all 3 dimensions (sagittal, coronal, and axial) of the orbitofacial defect as a preliminary step in the formation of a biomodel.

Graphic Jump Location
Place holder to copy figure label and caption
Figure 2.

Photographs of a 3-dimensional (3-D) computed tomographic biomodel with a polymethyl methacrylate (PMMA) implant. A, Front view of the biomodel (shown in violet) with the PMMA implant (shown in green) covering the orbital defect. B, Diagonal view of the same biomodel and implant demonstrating the 3-D effect and showing depth and dimension of the implant, which is fitted into the defect. The prefabricated PMMA implant (front [a] and diagonal [b] view) is constructed using computer-generated data and the biomodel. The generation of the 3-D biomodel is crucial, as it determines the exact shape and dimension of the future prefabricated PMMA implant.

Graphic Jump Location
Place holder to copy figure label and caption
Figure 3.

Intraoperative photographs showing accuate placement of the prefabricated polymethyl methacrylate (PMMA) implant. A, Formation of a thin meniscus of fluid between the bone implant interface is seen, which confirms proper placement of the PMMA implant over the anatomical defect. B, Clear distribution of fluid is seen over the bone under the transparent PMMA implant.

Graphic Jump Location
Place holder to copy figure label and caption
Figure 4.

Clinical features of the 9 patients who underwent surgical reconstruction of traumatic orbitofacial defects using polymethyl methacrylate implants. “Others” includes traumatic mydriasis, aphakia, retinopathy with optic neuropathy, and hypoesthesia over the frontal nerve region. VA indicates visual acuity; SS, superior sulcus.

Graphic Jump Location
Place holder to copy figure label and caption
Figure 5.

Photograph demonstrating significant posttraumatic enophthalmos, superior sulcus deformity, and cicatricial orbitopathy. This patient underwent high-resolution computed tomographic biomodeling, construction of a prefabricated polymethyl methacrylate implant, and reconstructive surgery with implantation.

Graphic Jump Location
Place holder to copy figure label and caption
Figure 6.

Postoperative photograph showing significant correction of the superior sulcus deformity, hypoglobus, and resolving enophthalmos. The patient was followed up for 10 years and showed no signs of any complication.

Graphic Jump Location
Place holder to copy figure label and caption
Figure 7.

Photographs of a patient with severe facial and orbital trauma after a fall from a 50-ft height. A, Significant posttraumatic hypoglobus, restrictive strabismus, enophthalmos, and orbital volume defect of the right eye. The patient underwent multiple reconstructive procedures of the face and orbit but still experienced diplopia and facial asymmetry before undergoing reconstructive surgery with implantation of a prefabricated, custom polymethyl methacrylate implant. B, Early postoperative photograph showing minimal orbital inflammation, with sutures at the superior eyelid crease and lateral canthal angle after reconstructive surgery. C, At 1-year follow-up there is improvement of the hypoglobus, superior sulcus deformity, and enophthalmos.

Graphic Jump Location
Place holder to copy figure label and caption
Figure 8.

Photographs showing computed tomographic biomodeling of the patient's anatomical defects using polymethyl methacrylate (PMMA) implants placed over the defect. Two separate PMMA implants were constructed for superior and inferior bony defects (A). Prefabricated custom PMMA implants were designed for the superior orbital rim (B) and the orbital floor (C), with a lock-and-key fit to the anatomical structures (D).

Graphic Jump Location
Place holder to copy figure label and caption
Figure 9.

Intraoperative photographs showing placement of a custom polymethyl methacrylate implant for a superior orbital rim defect. A, A superior eyelid crease incision was used to expose the fracture site. B, Fixation of the polymethyl methacrylate implant with metallic screws.

Graphic Jump Location
Place holder to copy figure label and caption
Figure 10.

Intraoperative photographs of a patient who underwent reconstruction of an orbital floor defect after trauma. A, An inferior transconjunctival incision was made with subperiosteal dissection to achieve access to the bony defect. B, Placement of a custom polymethyl methacrylate implant in the orbital floor defect.

Graphic Jump Location
Place holder to copy figure label and caption
Figure 11.

Preoperative and postoperative photographs of a patient who underwent reconstruction using a polymethyl methacrylate implant for multiple bony injuries and eyelid lacerations after a motor vehicle crash. A, Preoperative photograph showing right hypoglobus and enophthalmos in addition to midface descent and hollowing of the malar region. B, Postoperative photograph at 8-month follow-up showing improvement of the hypoglobus and facial symmetry.

Graphic Jump Location
Place holder to copy figure label and caption
Figure 12.

Coronal computed tomographic scan of the face demonstrating a comminuted fracture of the right maxillary anterior wall with hemorrhagic opacification of the right maxillary sinus. Also seen is the fracture of the right orbital floor involving the anterior ramus.

Graphic Jump Location
Place holder to copy figure label and caption
Figure 13.

Photographs of the computed tomographic biomodel and custom prefabricated implant for reconstruction of an orbitofacial defect. A, Computed tomographic biomodel prepared for the patient with an orbital floor fracture. The polymethyl methacrylate implant is shown in the inset. B, Photograph showing accurate placement of the polymethyl methacrylate implant over the orbital floor defect.

Graphic Jump Location

Tables

Table Graphic Jump LocationTable 1. Characteristics of Patients Who Underwent Reconstructive Procedures Using Custom, Prefabricated PMMA Implants
Table Graphic Jump LocationTable 2. Surgical Data of Patients Who Underwent Reconstruction Using PMMA Implants*

References

Eufinger  HWehmoller  MHarders  AHeuser  L. Prostheses for the reconstruction of skull defects. Int J Oral Maxillofac Surg 1995;24104- 110
PubMed
Clearihue  WJGroth  MJNoel  JC. PMMA implant fabrication, utilizing the Photo-polymer Solid Ground Forming Process. J Facial Somato Prosthet 1998;4105- 118
Taub  PJRudkin  GHClearihue  WJ  IIIMiller  TA. Prefabricated alloplastic implants for cranial defects. Plast Reconstr Surg 2003;1111233- 1240
PubMed
Janecka  IP New reconstructive technologies in skull base surgery: role of titanium mesh and porous polyethylene. Arch Otolaryngol Head Neck Surg 2000;126396- 401
PubMed
Steinhauser  EWHardt  N Secondary reconstruction of cranial defects. J Maxillofac Surg 1977;5192- 198
PubMed
Chiarini  LFigurelli  SPollastri  G  et al.  Cranioplasty using acrylic material: a new technical procedure. J Craniomaxillofac Surg 2004;325- 9
PubMed
Jordan  DROnge  PSAnderson  RLPatrinely  JRNerad  JA. Complications associated with alloplastic implants used in orbital fracture repair. Ophthalmology 1992;991600- 1608
PubMed
Rubin  PABilyk  JRShore  JW. Orbital reconstruction using porous polyethylene sheets. Ophthalmology 1994;1011697- 1708
PubMed
McNab  AA Sino-orbital fistula: two case reports. Clin Experiment Ophthalmol 2000;28324- 325
PubMed
Kinnunen  IAitasalo  KPollonen  MVarpula  M. Reconstruction of orbital floor fractures using bioactive glass. J Craniomaxillofac Surg 2000;28229- 234
PubMed
Ng  SGMadill  SAInkster  CFMaloof  AJLeatherbarrow  B. Medpor porous polyethylene implants in orbital blowout fracture repair. Eye 2001;15578- 582
PubMed
Mauriello  JA  JrHargrave  SYee  SMostafavi  RKapila  R Infection after insertion of alloplastic orbital floor implants. Am J Ophthalmol 1994;117246- 252
PubMed
Rosen  HM Porous, block hydroxyapatite as an interpositional bone graft substitute in orthognathic surgery. Plast Reconstr Surg 1989;83985- 990
PubMed
Merritt  KShafer  JWBrown  S Implant site infection rates with porous and dense materials. J Biomed Mater Res 1979;13101- 108
PubMed
Rosen  HM The response of porous hydroxyapatite to contiguous tissue infection. Plast Reconstr Surg 1991;881076- 1080
PubMed
Luparello  DBruschi  SVerna  G  et al.  Cranioplasty with polymethylmethacrylate: the clinico-statistical considerations. Minerva Chir 1998;53575- 579
PubMed

Correspondence

CME
Meets CME requirements for:
Browse CME for all U.S. States
Accreditation Information
The American Medical Association is accredited by the Accreditation Council for Continuing Medical Education to provide continuing medical education for physicians. The AMA designates this journal-based CME activity for a maximum of 1 AMA PRA Category 1 CreditTM per course. Physicians should claim only the credit commensurate with the extent of their participation in the activity. Physicians who complete the CME course and score at least 80% correct on the quiz are eligible for AMA PRA Category 1 CreditTM.
Note: You must get at least of the answers correct to pass this quiz.
You have not filled in all the answers to complete this quiz
The following questions were not answered:
Sorry, you have unsuccessfully completed this CME quiz with a score of
The following questions were not answered correctly:
Commitment to Change (optional):
Indicate what change(s) you will implement in your practice, if any, based on this CME course.
Your quiz results:
The filled radio buttons indicate your responses. The preferred responses are highlighted
For CME Course: A Proposed Model for Initial Assessment and Management of Acute Heart Failure Syndromes
Indicate what changes(s) you will implement in your practice, if any, based on this CME course.
NOTE:
Citing articles are presented as examples only. In non-demo SCM6 implementation, integration with CrossRef’s "Cited By" API will populate this tab (http://www.crossref.org/citedby.html).
Submit a Comment

Multimedia

Some tools below are only available to our subscribers or users with an online account.

Web of Science® Times Cited: 7

Related Content

Customize your page view by dragging & repositioning the boxes below.

Articles Related By Topic
Related Topics
PubMed Articles