Original Article |

Clinical Pilot Study of Intense Ultrasound Therapy to Deep Dermal Facial Skin and Subcutaneous Tissues FREE

Richard E. Gliklich, MD; W. Matthew White, MD; Michael H. Slayton, PhD; Peter G. Barthe, PhD; Inder Raj S. Makin, MD, PhD
[+] Author Affiliations

Correspondence: Richard E. Gliklich, MD, Division of Facial Plastic and Reconstructive Surgery, Department of Otolaryngology, Massachusetts Eye and Ear Infirmary, 243 Charles St, Boston, MA 02114 (Richard_Gliklich@meei.harvard.edu).


Author Affiliations: Division of Facial Plastic and Reconstructive Surgery, Department of Otology and Laryngology, Massachusetts Eye and Ear Infirmary, Harvard Medical School, Boston (Drs Gliklich and White); and Ulthera Inc, Mesa, Ariz (Drs Slayton, Barthe, and Makin).


Arch Facial Plast Surg. 2007;9(2):88-95. doi:10.1001/archfaci.9.2.88.
Text Size: A A A
Published online

Objective  To evaluate the clinical safety of intense ultrasound in the treatment of the dermis and subcutaneous tissues of the face and neck in terms of skin inflammation, pain, adverse events, and histologic features.

Design  In an open-label, phase 1 study, patients scheduled to undergo a rhytidectomy were enrolled into immediate (face-lift surgery within 24 hours of intense ultrasound treatment) and delayed (face-lift surgery 4-12 weeks after treatment) treatment groups. Intense ultrasound treatments were performed as a series of several linear exposures delivered 1.5 to 2.0 mm apart with the use of 1 of 3 available handpieces with different focal depths. Subject pain ratings and standardized digital photographs were obtained at uniform points. Photographs were blindly rated for inflammation. Histologic evaluation of treated tissues was performed with nitroblue tetrazolium chloride viability stain.

Results  Fifteen subjects with a mean ± SD age of 53 ± 7 years were enrolled. Seven subjects were nonrandomly assigned to the immediate group and 8 were in the delayed group. On histologic examination, thermal injury zones were consistently identified in the dermis at exposure levels greater than 0.5 J as focal areas of denatured collagen. At this threshold level or above, most patient exposures were associated with transient superficial skin erythema and slight to mild discomfort on a standardized pain scale. No other adverse effects were noted in any case. Thermal injury zones were produced in the expected linear pattern and were consistent in size and depth from zone to zone. Increasing source power did not increase the depth of the epicenter of the thermal injury zone. Epidermis was spared in all cases.

Conclusion  In this first clinical study of intense ultrasound therapy to facial tissues, the intense ultrasound system allowed for the safe and well-tolerated placement of targeted, precise, and consistent thermal injury zones in the dermis and subcutaneous tissues with sparing of the epidermis.

Figures in this Article

Various noninvasive devices have been developed in an effort to treat facial rhytids.1 These devices can be grossly categorized as either ablative skin resurfacing (ASR) or nonablative skin rejuvenation (NSR). The ASR modalities (microdermabrasion, chemical peels, and carbon dioxide and erbium-YAG laser therapy) work by removing the superficial layer of the epidermis and inducing thermal injury to the dermis. All these ASR modalities have demonstrated superior efficacy in treating superficial rhytids in the aging face. However, because of the removal of the epidermis, patients treated with ASR can have prolonged erythema, infections, and permanent pigmentary changes.

The NSR devices (intense pulsed light and light-emitting diode, radiofrequency, Nd:YAG, and pulsed dye lasers) have been designed to selectively induce thermal injury within the dermis while sparing the overlying epidermis.2-3 This controlled dermal heating incites a “wound-healing” response through the liberation of several cytokines that stimulate fibroblasts to synthesize and lay down new collagen. The lack of epidermal damage with NSR techniques has significantly reduced the treatment-related side effects seen with ASR, yet the overall treatment efficacy, especially with superficial rhytids, has been suboptimal when compared with ASR.4-6

Intense ultrasound (IUS) is an energy modality that can propagate through tissues, resulting in selective thermal coagulative change within the focal region of the beam while leaving the remaining regions unaffected.7 Ultrasound waves induce a vibration in the composite molecules of a given tissue, and the friction developed between the molecules is the source of the generated heat. It has been well established in the literature that IUS fields transcutaneously directed into whole-organ soft tissue can produce coagulative necrosis resulting primarily from thermal mechanisms.8-10 During the past decade, the clinical use of focused IUS has been investigated as a noninvasive surgical tool to treat whole-organ tumors, such as liver, breast, and uterus. To our knowledge, this technology has not been previously applied to human facial tissue.

Working in conjunction with the manufacturer (Ulthera Inc, Mesa, Ariz), we have developed a novel IUS device capable of creating thermal injury zones (TIZ) at depths up to 6 mm. This device is able to focus energy within tissue to produce a 25-mm line of discrete TIZs spaced 0.5 to 5.0 mm apart. Furthermore, both imaging and targeted energy exposure can be accomplished with the same handpiece. Preclinical studies conducted in our department, with human cadaveric facial tissues, have demonstrated that the IUS system can reliably create small, well-defined TIZs in the subdermal soft tissue and deeper superficial musculoaponeurotic system layers while preserving immediately adjacent soft tissue and structures.8, 11

In summary, the IUS system is a novel device designed to target and deliver focused energy to a specific skin layer. The purpose of this study was to evaluate the clinical safety of the IUS in the treatment of the face and neck, using skin inflammation, patient-reported pain, and histologic evaluation of injury as the primary end points after IUS exposure over a range of energy levels and source conditions.

STUDY DESIGN

We conducted an open-label, prospective, pilot phase 1 human study to determine skin tissue response after exposure to an investigational ultrasound device (the IUS system). This study was approved by the Massachusetts Eye and Ear Infirmary Institutional Review Board for Human Studies.

Adult patients scheduled to undergo a limited rhytidectomy (mini–face-lift) procedure who provided informed consent were eligible for the study. Exclusion criteria included the following: active systemic or local infections; systemic or local skin disease; previous facial surgery (including rhytidectomy, lesion removal, laser or other skin resurfacing, radiofrequency treatments, and injections of filler materials); scarring of any etiology in the planned treatment areas; psychiatric illness; and inability to provide informed consent.

Subjects were assigned by preference for timing of their rhytidectomy into 1 of 2 groups: (1) patients who would receive a surgical mini–face-lift within 24 hours after IUS exposures (immediate group), and (2) patients who would receive a mini–face-lift 4 to 12 weeks after IUS exposures (delayed group).

IUS DEVICE

The prototype IUS device used in this study (UltraSite GT; Ulthera Inc) was specifically designed for facial soft tissues to create TIZs at certain depths with specified patterns. This device consists of a central power unit, a computer, and a delivery handpiece (Figure 1). The same handpiece enables sequential imaging (to evaluate layers and structures and to target treatment) and delivery of the ultrasound energy. The console allows for multiple source settings to be controlled, including power output, exposure time, length of exposure line, distance between exposure zones, and time delay after each exposure.

Place holder to copy figure label and caption
Figure 1.

Intense ultrasound system. Imaging and treatment are possible with the same handpiece.

Graphic Jump Location

Previous studies performed in our laboratory in porcine and human cadaveric tissues have demonstrated that higher–ultrasound-frequency handpieces have shallower focal depths (ie, produce TIZs more superficial in tissue), while lower-frequency handpieces have deeper focal depths (ie, produce TIZs at greater depths).8 Three handpieces were created specifically for this experiment: (1) superficial: 7.5 MHz, 3.0-mm focus depth; (2) intermediate: 7.5 MHz, 4.5-mm focus depth; and (3) deep: 4.4 MHz, 4.5-mm focus depth.

On the basis of our previous work in human cadaveric tissues, ultrasound exposures were administered in this study starting with the lowest ultrasound exposure (0.5 J) seen to induce TIZs in human cadaveric facial tissues by means of several combinations of power and exposure durations.

TREATMENT PROCEDURE

The surgeon (R.E.G.) first estimated the amount of periauricular skin area that was destined to be excised during the patient's limited rhytidectomy. This area was demarcated with the use of a surgical marker. Topical anesthetic cream (2.5% lidocaine and 2.5% prilocaine; EMLA; AstraZeneca, Wilmington, Del) was applied to the treatment area and left under occlusion for approximately 45 minutes.

Within each skin area to be excised during face-lift surgery, a series of paired microtattoos were placed 10 mm apart with sterile india ink to indelibly mark the positions of each of the 5 to 10 ultrasound exposure lines. The size of the skin to be excised was unique for each patient and, in some cases, limited the number of ultrasound exposure sites. The microtattoos were used to correlate the position of surface exposures to histologic examination of the tissue.

After cleansing the skin surface with an alcohol swab, ultrasound gel (Parker Laboratories, Fairfield, NJ) was applied to the skin. Each proposed area of treatment was then imaged (Figure 2) and the diagnostic image was stored. The IUS handpiece was positioned by alignment with a skin microtattoo pair, a diagnostic image was captured, and the treatment mode was then activated. The treating component of the handpiece mechanically moves in a straight line, at the same source conditions (power, duration) in accordance with the selected treatment variables (length of treatment, spacing of exposures) for distances up to 15 mm, generally producing 5 to 7 uniform tissue exposures per each horizontal “line” of IUS treatment. Several IUS treatments were performed according to the available tissue in each patient (ie, tissue planned for subsequent excision). This allowed the comparison of multiple source settings from each side of the patient's face after treatment on both sides.

Place holder to copy figure label and caption
Figure 2.

Ultrasound image of the left preauricular region in patient 8. Imaging allows identification of tissue layers (epidermis, dermis, subcutaneous fat, and superficial musculoaponeurotic system [SMAS]). This assessment can help to guide source settings and ensure that the handpiece is coupled to the tissue. Also, the selected length of the treatment line and pulse spacing is displayed onscreen.

Graphic Jump Location

High-resolution digital photography of the periauricular skin areas was performed with a digital single-lens-reflex camera (Nikon D-70; Nikon USA, Melville, NY). Standardized conditions for distance (same f-stop) and lighting were used for photography. Photographs were taken of both the left- and right-side profiles of the face at the following time points: at the screening visit, after microtattoo placement just before IUS treatment, immediately after IUS treatment, 10 minutes after IUS treatment, 48 hours' follow-up (in the delayed group), and just before the surgical mini–face-lift. With the use of image processing software (NIH Image J; http://rsbweb.nih.gov/ij/), postexposure digital photographs were compared with the pretreatment images.

PAIN, INFLAMMATION, AND ADVERSE EVENTS

After each individual IUS exposure line, the subject was asked to rate sensation by using a 5-point scale as follows: no sensation (0); warm (1); hot, with mild transient pain (2); hot, with moderate transient pain (3); hot, with strong but transient pain (4); and hot, with strong and persistent pain (5). For dermal exposures, the gross cutaneous responses were blindly graded by a single rater (W.M.W.) from photographs taken immediately (papule formation) and 10 minutes (erythema or edema) after exposure on a scale of 0 to 3 as follows: absent (0), slight (1), moderate (2), and prominent (3). All patients were asked to report whether they had experienced any adverse events of any type at each visit.

HISTOLOGIC ANALYSIS

After surgical excision of tissue, all treatment sites were oriented in the anatomic position and placed in a freezer (−15°C). The frozen tissue was sectioned perpendicular to the horizontal IUS exposure lines and immersed in nitroblue tetrazolium chloride (NBTC) solution for 24 hours. The remaining tissue was embedded and submitted for histopathologic examination.

Thermal lesions within collagen were evaluated by the NBTC viability stain,12 a frequently used stain to evaluate laser-induced thermal injury. Blue staining of cells on frozen section with NBTC confirms viability, and the absence of blue staining is indicative of an area of coagulative necrosis. A positive histologic finding of coagulative change is recorded when at least 3 consecutive histologic sections indicate a region of thermal necrosis. In this manner, the 3-dimensional nature of the thermal coagulative zones was confirmed.

Fifteen subjects were assigned to 1 of 2 treatment groups (Table). There were 2 males and 13 females. Mean ± SD age was 53 ± 7 years. Seven patients underwent face-lift within 24 hours after IUS treatment (immediate group) with complete excision of the treated areas. Eight patients underwent face-lift surgery between 4 and 12 weeks (mean, 5.5 weeks) after IUS treatment (delayed group). More than 1400 individual IUS exposures were administered and evaluated in the patients.

SENSORY AND SKIN RESPONSE

As shown in the Table, patients varied in their pain response to the IUS treatments. All patients were pretreated with topical lidocaine. Only 1 patient required injection with local anesthetic to complete the treatments, mostly secondary to anxiety. Most patients rated the pain sensation from 0 (no sensation) to 3 (hot, with moderate transient pain) immediately after treatment, and pain was rated as 0 (no sensation) in all patients by 10 minutes after treatment. All patients in the delayed group reported no persistent pain at 48 hours after treatment.

Most patients in both immediate and delayed treatment groups demonstrated transient, mild erythema immediately after treatment. Figure 3 demonstrates clinical photographs taken in a patient treated with the most superficial handpiece (7.5 MHz/3 mm). Here we can see persistent erythema at the 10-minute point (rated 1, slight). Erythema generally dissipated by 10 minutes in all patients treated with the superficial handpiece. In patients treated with the deeper focal probes (7.5 MHz/4.5 mm [Figure 4] and 4.4 MHz/4.5 mm [Figure 5]), a milder inflammatory response was seen. No persistent erythema was discernible at 48 hours in any patient. In 2 patients treated with the superficial 3.0-mm focal-depth handpiece at the highest energy setting (1.0 J), a linear inflammatory wheal appeared immediately after treatment but completely resolved by 48 hours. Clinically, no signs of epidermal disruption or slough (Figures 3, 4, and 5) or any additional delayed adverse effects (eg, persistent pain or inflammation) were seen in any patient.

Place holder to copy figure label and caption
Figure 3.

Patient 11. Clinical photographs taken before treatment with a 7.5-MHz/3-mm handpiece (A) and 10 minutes after exposure (B). Nine lines of exposure were placed at 0.625 J. Inflammation score was 1; sensation, 2. The scale is in millimeters.

Graphic Jump Location
Place holder to copy figure label and caption
Figure 4.

Patient 13. Clinical photographs taken before treatment with a 7.5-MHz/4.5-mm handpiece (A) and 10 minutes after exposure (B). Ten lines of exposure were placed at 1.05 J. Inflammation score was 0; sensation, 0. The scale is in millimeters.

Graphic Jump Location
Place holder to copy figure label and caption
Figure 5.

Patient 12. Clinical photographs taken before treatment with a 4.4-MHz/4.5-mm handpiece (A) and 10 minutes after exposure (B). Eight lines of exposure were placed at 1.6 J. Inflammation score was 0; sensation, 1 to 2. The scale is in millimeters.

Graphic Jump Location

Figures 6, 7, and 8 summarize the sensory and inflammatory responses for all of the exposures with the 3 handpiece types used in the study. Each bar demonstrates average patient response, as multiple unique exposure lines were made at the same source power-time (energy) combination. For each graph (ie, each handpiece type), the data could be divided into 3 categories based on the power settings (watts) used for the IUS exposures. In each graph, the first grouping of data bars represents the higher source power (watts) settings, or “high power.” The second grouping of data bars represents lower source power (watts) settings, or “low power.” The specific power levels for each given handpiece are given in the figure legends.

Place holder to copy figure label and caption
Figure 6.

Patient responses to treatment with a 7.5-MHz/3.0-mm handpiece. Group 1 had high power levels (30-45 W at 0.30-1.13 J). In group 2 the power levels were kept lower (15-25 W at 0.15-1.00 J). Group 3 had dense pattern exposures (25 W, 0.625 J).

Graphic Jump Location
Place holder to copy figure label and caption
Figure 7.

Patient responses to treatment with a 7.5-MHz/4.5-mm handpiece. Group 1 had higher power levels (50-60 W at 0.50-1.93 J). In group 2 the power levels were kept lower (30-35 W at 0.90-1.50 J). Group 3 had dense pattern exposures (30 W, 1.05 J).

Graphic Jump Location
Place holder to copy figure label and caption
Figure 8.

Patient responses to treatment with a 4.4-MHz/4.5-mm handpiece. Group 1 had higher power levels (60-70 W at 0.6-2.1 J). In group 2 the power levels were kept lower (40-50 W at 0.50-1.60 J). Group 3 had dense pattern exposures (40 W, 1.60 J).

Graphic Jump Location

From the individual graphs, it is evident that the high-power settings resulted nominally in a greater sensory response, whereas the low-power source conditions were better tolerated by the subjects. Furthermore, when the 3 unique handpieces were compared over similar source conditions, the superficial handpiece (7.5 MHz/3 mm [Figure 6]) produced less of a sensory response than did the intermediate and deep handpieces, both of which were focused at 4.5 mm (Figures 7 and 8). However, higher-frequency handpieces (7.5 MHz [Figures 6 and 7]) tended to produce a greater inflammatory response than did the lower-frequency handpiece (4.4 MHz [Figure 8]). The average inflammatory response for all of the exposures made with the 3 IUS handpiece types did not exceed 2 (moderate inflammation) on a 0 to 3 ordinal scale.

The third grouping of data bars seen in the graphs (Figures 6-8, rightmost bars) consists of the average sensory and inflammation scores of patients after placement of a higher density of parallel IUS exposures. In this higher-density pattern, multiple consecutive parallel exposure lines were delivered at the same unique power setting, such that an array of densely spaced TIZs were deposited per unit volume of tissue. This was in contrast to the previous low-power and high-power groups, in which individual exposure lines were delivered at different source settings. These source settings chosen for the denser pattern were from the low-power source setting groups for each handpiece. Again, when denselyspaced TIZs were deposited in a higher-density pattern, no clinically significant sensory discomfort or inflammation was noted (Figures 3-5).

NBTC VITAL STAINING

On gross and histologic examination, TIZs were consistently identified in the dermis (corresponding to the focal point of the exposures) as areas of collagen denaturation at exposure power levels greater than 0.5 J. At this TIZ threshold level or above, most patient exposures were associated with transient superficial skin erythema and slight to mild discomfort (average pain sensation score, 3 [hot, with moderate, transient pain]). The TIZs were reproducibly created in the expected linear pattern and were consistent from zone to zone (Figure 9). The depth of the TIZs was consistent from spot to spot for the same handpiece. Increasing source power did not increase the depth of the epicenter of the TIZ (Figures 10, 11, and 12). The epidermis was spared in all cases.

Place holder to copy figure label and caption
Figure 9.

Nitroblue tetrazolium chloride (NBTC) staining of grossly excised preauricular tissue, sectioned perpendicularly to the lines of treatment. Two thermal injury zones can be seen by their loss of blue NBTC vital staining just below the skin surface (arrows). The treatment was performed with a 7.5-MHz/3-mm handpiece on patient 8 (scale, 1 mm per division). The scale is in millimeters.

Graphic Jump Location
Place holder to copy figure label and caption
Figure 10.

Digital photomicrograph of a frozen section stained with nitroblue tetrazolium chloride (NBTC) (eosin counterstain). A thermal injury zone is shown by the loss of blue NBTC vital stain and dilated collagen fibers just below the skin epidermis (dotted line). Note the complete epidermal preservation. The treatment was performed with a 7.5-MHz/3-mm handpiece (0.9 J) on patient 8 (original magnification, ×20).

Graphic Jump Location
Place holder to copy figure label and caption
Figure 11.

Digital photomicrograph of a frozen section stained with nitroblue tetrazolium chloride (NBTC) (eosin counterstain). A thermal injury zone is shown by the loss of blue NBTC vital stain and dilated collagen fibers just below the skin epidermis (dotted line). The treatment was performed with a 7.5-MHz/4.5-mm handpiece (1.2 J) on patient 8 (original magnification, ×20).

Graphic Jump Location
Place holder to copy figure label and caption
Figure 12.

Digital photomicrograph of a frozen section stained with nitroblue tetrazolium chloride (NBTC) (eosin counterstain). A thermal injury zone is shown by the loss of blue NBTC vital stain and dilated collagen fibers just below the skin epidermis (dotted line). Note the epidermal preservation. The treatment was performed with a 4.4-MHz/4.5-mm handpiece (1.6 J) on patient 12 (original magnification, ×20).

Graphic Jump Location

This open-label pilot clinical trial was conducted to confirm the safety of IUS treatment to facial dermis and subcutaneous tissue. The results demonstrate that, within the range of source conditions (0.5-2.1 J) used for the clinical study, the IUS exposures were well tolerated by the subjects. Furthermore, on the exposed site, there was no detrimental change in the epidermis (Figures 3-5), as well as no delayed sequelae in the treated tissue (4-12 weeks posttreatment). During this 15-patient study, there were no findings related to a detrimental impact of the IUS exposures on the facial nerve or its branches. This negative finding was anticipated on the basis of the safe use of other energy-based cosmetic devices on which the IUS device is predicated, as well as the deeper location of the facial nerve in the periauricular region13 compared with the depth at which the IUS system is expected to cause a tissue effect.

The purpose of most energy-delivery devices in cosmetic applications is to achieve a controlled thermal injury in the skin tissue. This damage results in an immediate contraction of the collagen tissue and initiates a repair process in which a new collagen matrix is laid down subepidermally.1, 6, 14 The IUS device is designed to achieve selective regions (on the order of 1 mm3 per individual exposure shot) of thermal coagulative change below the epidermis while sparing the epidermis, adding to the overall safety of this technique. An aim of this study was, therefore, to confirm the presence of selective regions of acute thermal microablative change resulting from exposures from various IUS handpiece types at particular source conditions. Within the range of source conditions (power-time combinations, frequency, etc) used in this study, selective thermal coagulative changes were confirmed with each of the handpiece types used (Table). The coagulative damage was identified histopathologically, by the classic lack of cellular organization in the ultrasound-exposed region, as well as the denaturation of collagen in the region. These findings are evident by the lack of vital stain (NBTC) uptake (Figures 10-12) in the thermally affected region. These histologic changes are well documented in the literature as being a result of thermal damage to the skin tissue.6, 8, 12, 15 The positive findings were identified primarily in the tissue excised from the patient group undergoing an immediate mini–face-lift surgery. Once again, the epidermis was spared in all cases in this study, and the region of thermal microablative damage from the IUS exposure was particularly selective (on the order of 1 mm3).

No definitive and significant histologic change could be identified in the evaluation of tissue from the delayed patient group. The lack of significant histologic findings in the tissue from the delayed group suggests resolution of the microablative injury that was caused by the IUS exposures. A limitation of this study was the lack of electron microscopy to further evaluate collagen changes with both injury and healing.16

This clinical study of IUS to the face and neck demonstrated that IUS can produce targeted, consistent, and reproducible thermal injury zones in the dermis and subcutaneous tissues. The ability to both target and noninvasively treat facial tissues is unique to the ultrasound device. Treatments were associated with limited transient erythema, mild discomfort, and histologic evidence of collagen denaturation in the dermis with sparing of the overlying epidermal layer. Further studies are needed to determine the efficacy of this new treatment modality, which has important implications for facial aesthetic procedures.

Correspondence: Richard E. Gliklich, MD, Division of Facial Plastic and Reconstructive Surgery, Department of Otolaryngology, Massachusetts Eye and Ear Infirmary, 243 Charles St, Boston, MA 02114 (Richard_Gliklich@meei.harvard.edu).

Accepted for Publication: October 8, 2006.

Author Contributions:Study concept and design: Gliklich, White, Slayton, Barthe, and Makin. Acquisition of data: Gliklich, White, and Makin. Analysis and interpretation of data: Gliklich, White, Slayton, Barthe, and Makin. Drafting of the manuscript: Gliklich, White, and Makin. Critical revision of the manuscript for important intellectual content: Gliklich, White, Slayton, Barthe, and Makin. Statistical analysis: Gliklich, White, and Makin. Obtained funding: Gliklich and White. Administrative, technical, and material support: Gliklich, White, Slayton, Barthe, and Makin. Study supervision: Gliklich and White.

Financial Disclosure: Drs Gliklich and White have received a grant or funding from Ulthera Inc.

Funding/Support: This study was supported in part by Ulthera Inc.

Hruza  GJ Rejuvenating the aging face Arch Facial Plast Surg 2004;6366- 369
PubMed Link to Article
Kim  KHGeronemus  RG Nonablative laser and light therapies for skin rejuvenation Arch Facial Plast Surg 2004;6398- 409
PubMed Link to Article
Dierickx  CCAnderson  RR Visible light treatment of photoaging Dermatol Ther 2005;18191- 208
PubMed Link to Article
Kelly  KMMajaron  BNelson  JS Nonablative laser and light rejuvenation Arch Facial Plast Surg 2001;3230- 234
PubMed Link to Article
Grema  HGreve  BRaulin  C Facial rhytids: subsurfacing or resurfacing Lasers Surg Med 2003;32405- 412
PubMed Link to Article
Manstein  DHerron  GSSink  RKTanner  HAnderson  RR Fractional photothermolysis: a new concept for cutaneous remodeling using microscopic patterns of thermal injury Lasers Surg Med 2004;34426- 438
PubMed Link to Article
Kennedy  JEter Haar  GRCranston  D High intensity focused ultrasound: surgery of the future Br J Radiol 2003;76590- 599
PubMed Link to Article
White  WMLaubach  HJMakin  IRSBarthe  PGSlayton  MHGliklich  RE Selective transcutaneous delivery of energy to facial subdermal tissues using the ultrasound therapy system [abstract] Lasers Surg Med 2006;38suppl 18113
Laubach  HJBarthe  PGMakin  IRSSlayton  MHManstein  DConfined thermal damage with intense ultrasound (IUS) [abstract]Lasers Surg Med 2006;38suppl 1832
Link to Article
Makin  IRMast  TDFaidi  WRunk  MMBarthe  PGSlayton  MH Miniaturized ultrasound arrays for interstitial ablation and imaging Ultrasound Med Biol 2005;311539- 1550
PubMed Link to Article
White  WMMakin  IRBarthe  PGSlayton  MHGliklich  RE Selective creation of thermal injury zones in the superficial musculoaponeurotic system using intense ultrasound therapy: a new target for noninvasive facial rejuvenation Arch Facial Plast Surg 2007;922- 29
PubMed Link to Article
Neumann  RAKnobler  RMPieczkowski  FGebhart  W Enzyme histochemical analysis of cell viability after argon laser-induced coagulation necrosis of the skin J Am Acad Dermatol 1991;25991- 998
PubMed Link to Article
Rudolph  R Depth of the facial nerve in face lift dissections Plast Reconstr Surg 1990;85537- 544
PubMed Link to Article
Fitzpatrick  RERostan  EFMarchell  N Collagen tightening induced by carbon dioxide laser versus erbium:YAG laser Lasers Surg Med 2000;27395- 403
PubMed Link to Article
Laubach  H-JBarthe  PGMakin  IRSSlayton  MHManstein  D Confined thermal damage with intense ultrasound (IUS) [abstract] Lasers Surg Med 2006;38suppl 1828
Link to Article
Sadick  NSTrelles  MA Nonablative wrinkle treatment of the face and neck using a combined diode laser and radiofrequency technology Dermatol Surg 2005;311695- 1699
PubMed Link to Article

Figures

Place holder to copy figure label and caption
Figure 1.

Intense ultrasound system. Imaging and treatment are possible with the same handpiece.

Graphic Jump Location
Place holder to copy figure label and caption
Figure 2.

Ultrasound image of the left preauricular region in patient 8. Imaging allows identification of tissue layers (epidermis, dermis, subcutaneous fat, and superficial musculoaponeurotic system [SMAS]). This assessment can help to guide source settings and ensure that the handpiece is coupled to the tissue. Also, the selected length of the treatment line and pulse spacing is displayed onscreen.

Graphic Jump Location
Place holder to copy figure label and caption
Figure 3.

Patient 11. Clinical photographs taken before treatment with a 7.5-MHz/3-mm handpiece (A) and 10 minutes after exposure (B). Nine lines of exposure were placed at 0.625 J. Inflammation score was 1; sensation, 2. The scale is in millimeters.

Graphic Jump Location
Place holder to copy figure label and caption
Figure 4.

Patient 13. Clinical photographs taken before treatment with a 7.5-MHz/4.5-mm handpiece (A) and 10 minutes after exposure (B). Ten lines of exposure were placed at 1.05 J. Inflammation score was 0; sensation, 0. The scale is in millimeters.

Graphic Jump Location
Place holder to copy figure label and caption
Figure 5.

Patient 12. Clinical photographs taken before treatment with a 4.4-MHz/4.5-mm handpiece (A) and 10 minutes after exposure (B). Eight lines of exposure were placed at 1.6 J. Inflammation score was 0; sensation, 1 to 2. The scale is in millimeters.

Graphic Jump Location
Place holder to copy figure label and caption
Figure 6.

Patient responses to treatment with a 7.5-MHz/3.0-mm handpiece. Group 1 had high power levels (30-45 W at 0.30-1.13 J). In group 2 the power levels were kept lower (15-25 W at 0.15-1.00 J). Group 3 had dense pattern exposures (25 W, 0.625 J).

Graphic Jump Location
Place holder to copy figure label and caption
Figure 7.

Patient responses to treatment with a 7.5-MHz/4.5-mm handpiece. Group 1 had higher power levels (50-60 W at 0.50-1.93 J). In group 2 the power levels were kept lower (30-35 W at 0.90-1.50 J). Group 3 had dense pattern exposures (30 W, 1.05 J).

Graphic Jump Location
Place holder to copy figure label and caption
Figure 8.

Patient responses to treatment with a 4.4-MHz/4.5-mm handpiece. Group 1 had higher power levels (60-70 W at 0.6-2.1 J). In group 2 the power levels were kept lower (40-50 W at 0.50-1.60 J). Group 3 had dense pattern exposures (40 W, 1.60 J).

Graphic Jump Location
Place holder to copy figure label and caption
Figure 9.

Nitroblue tetrazolium chloride (NBTC) staining of grossly excised preauricular tissue, sectioned perpendicularly to the lines of treatment. Two thermal injury zones can be seen by their loss of blue NBTC vital staining just below the skin surface (arrows). The treatment was performed with a 7.5-MHz/3-mm handpiece on patient 8 (scale, 1 mm per division). The scale is in millimeters.

Graphic Jump Location
Place holder to copy figure label and caption
Figure 10.

Digital photomicrograph of a frozen section stained with nitroblue tetrazolium chloride (NBTC) (eosin counterstain). A thermal injury zone is shown by the loss of blue NBTC vital stain and dilated collagen fibers just below the skin epidermis (dotted line). Note the complete epidermal preservation. The treatment was performed with a 7.5-MHz/3-mm handpiece (0.9 J) on patient 8 (original magnification, ×20).

Graphic Jump Location
Place holder to copy figure label and caption
Figure 11.

Digital photomicrograph of a frozen section stained with nitroblue tetrazolium chloride (NBTC) (eosin counterstain). A thermal injury zone is shown by the loss of blue NBTC vital stain and dilated collagen fibers just below the skin epidermis (dotted line). The treatment was performed with a 7.5-MHz/4.5-mm handpiece (1.2 J) on patient 8 (original magnification, ×20).

Graphic Jump Location
Place holder to copy figure label and caption
Figure 12.

Digital photomicrograph of a frozen section stained with nitroblue tetrazolium chloride (NBTC) (eosin counterstain). A thermal injury zone is shown by the loss of blue NBTC vital stain and dilated collagen fibers just below the skin epidermis (dotted line). Note the epidermal preservation. The treatment was performed with a 4.4-MHz/4.5-mm handpiece (1.6 J) on patient 12 (original magnification, ×20).

Graphic Jump Location

Tables

References

Hruza  GJ Rejuvenating the aging face Arch Facial Plast Surg 2004;6366- 369
PubMed Link to Article
Kim  KHGeronemus  RG Nonablative laser and light therapies for skin rejuvenation Arch Facial Plast Surg 2004;6398- 409
PubMed Link to Article
Dierickx  CCAnderson  RR Visible light treatment of photoaging Dermatol Ther 2005;18191- 208
PubMed Link to Article
Kelly  KMMajaron  BNelson  JS Nonablative laser and light rejuvenation Arch Facial Plast Surg 2001;3230- 234
PubMed Link to Article
Grema  HGreve  BRaulin  C Facial rhytids: subsurfacing or resurfacing Lasers Surg Med 2003;32405- 412
PubMed Link to Article
Manstein  DHerron  GSSink  RKTanner  HAnderson  RR Fractional photothermolysis: a new concept for cutaneous remodeling using microscopic patterns of thermal injury Lasers Surg Med 2004;34426- 438
PubMed Link to Article
Kennedy  JEter Haar  GRCranston  D High intensity focused ultrasound: surgery of the future Br J Radiol 2003;76590- 599
PubMed Link to Article
White  WMLaubach  HJMakin  IRSBarthe  PGSlayton  MHGliklich  RE Selective transcutaneous delivery of energy to facial subdermal tissues using the ultrasound therapy system [abstract] Lasers Surg Med 2006;38suppl 18113
Laubach  HJBarthe  PGMakin  IRSSlayton  MHManstein  DConfined thermal damage with intense ultrasound (IUS) [abstract]Lasers Surg Med 2006;38suppl 1832
Link to Article
Makin  IRMast  TDFaidi  WRunk  MMBarthe  PGSlayton  MH Miniaturized ultrasound arrays for interstitial ablation and imaging Ultrasound Med Biol 2005;311539- 1550
PubMed Link to Article
White  WMMakin  IRBarthe  PGSlayton  MHGliklich  RE Selective creation of thermal injury zones in the superficial musculoaponeurotic system using intense ultrasound therapy: a new target for noninvasive facial rejuvenation Arch Facial Plast Surg 2007;922- 29
PubMed Link to Article
Neumann  RAKnobler  RMPieczkowski  FGebhart  W Enzyme histochemical analysis of cell viability after argon laser-induced coagulation necrosis of the skin J Am Acad Dermatol 1991;25991- 998
PubMed Link to Article
Rudolph  R Depth of the facial nerve in face lift dissections Plast Reconstr Surg 1990;85537- 544
PubMed Link to Article
Fitzpatrick  RERostan  EFMarchell  N Collagen tightening induced by carbon dioxide laser versus erbium:YAG laser Lasers Surg Med 2000;27395- 403
PubMed Link to Article
Laubach  H-JBarthe  PGMakin  IRSSlayton  MHManstein  D Confined thermal damage with intense ultrasound (IUS) [abstract] Lasers Surg Med 2006;38suppl 1828
Link to Article
Sadick  NSTrelles  MA Nonablative wrinkle treatment of the face and neck using a combined diode laser and radiofrequency technology Dermatol Surg 2005;311695- 1699
PubMed Link to Article

Correspondence

CME
Meets CME requirements for:
Browse CME for all U.S. States
Accreditation Information
The American Medical Association is accredited by the Accreditation Council for Continuing Medical Education to provide continuing medical education for physicians. The AMA designates this journal-based CME activity for a maximum of 1 AMA PRA Category 1 CreditTM per course. Physicians should claim only the credit commensurate with the extent of their participation in the activity. Physicians who complete the CME course and score at least 80% correct on the quiz are eligible for AMA PRA Category 1 CreditTM.
Note: You must get at least of the answers correct to pass this quiz.
You have not filled in all the answers to complete this quiz
The following questions were not answered:
Sorry, you have unsuccessfully completed this CME quiz with a score of
The following questions were not answered correctly:
Commitment to Change (optional):
Indicate what change(s) you will implement in your practice, if any, based on this CME course.
Your quiz results:
The filled radio buttons indicate your responses. The preferred responses are highlighted
For CME Course: A Proposed Model for Initial Assessment and Management of Acute Heart Failure Syndromes
Indicate what changes(s) you will implement in your practice, if any, based on this CME course.